Welcome Guest from United States
Sign In Change Country
  0 Items
Search:

 DELTA V - The Widest Possible Range

 

The Widest Possible Range Available on Isotope Ratio MS

Universal Triple Collector
The universal triple collector is standard on the DELTA V platform and is suitable for all standard applications involving N2, CO, NO, O2, CO2, N2O and SO2.

D/H Collector
A H2 collector with two additional Faraday cups for m/z 2 and 3 is optional for both the DELTA V Advantage and the DELTA V Plus. Because measurement of D/H in Continuous Flow mode requires complete removal of low energy 4He+ ions from the HD+ ion beam at m/z 3, a unique ion optical element is built into the m/z 3 cup which removes all ions other than HD+, even at high He carrier gas loads.

Individual Collector Arrangements
The DELTA V Plus collector array is prepared for up to 10 collectors with deep individually shielded long-life Faraday cups. It offers the choice of several collector widths for optimal tailored solutions combining specific and standard cup systems within the same collector array. Configurations include e.g. m/z 28 to 46 for all major gases and isotopes in air; N2+ — NO+ — N2O+ for N2O isotopomers; simultaneous CO+ — CO2+ experiments; m/z 48 to 66 for SO+ — SO2+ for 18O-corrected SO2 measurements.

Gas Collector Arrangements for Masses (m/z)
H2 2   3
N2   28 29 30  
CO   28 29 30  
NO   30 31 32  
O2   32 33 34  
CO2   44 45 46  
N2O   44 45 46  
SO2   64 66  
DELTA V Plus - Additional Collector Arrangements
Air 28 29 32 33 34 36 40 44 45 46
N2O 28 29 30 31 32   44 45 46
CO, CO2 28 29 30   44 45 46
N2, CO2 28 29 30   44 45 46
SO2   64 65 66
SO, SO2   48 49 50 64 65 66
CH3Cl   50 51 52 53  
CH3Br   94 96
Ne 20 21 22  
Ar   36 38 40  
Kr   78 80 82 84 86

DELTA V Collector Unit