Welcome Guest from United States
Sign In Change Country
  0 Items
Search:

 DELTA V - Installation Requirements

 

Installation Requirements

DELTA V
Environment Heat output 1.5 kW during normal operation,
2.0 kW during bakeout (basic system).
Ambient temperature must be 18 - 28 °C
with a relative humidity of 20 - 70 %
Power 3 -phase 230/400 V -10 % + 6%,
fused 16 A / phase (single-phase optional),
50/60 Hz, dedicated line voltage
must be free of spikes.
Maximum permissible voltage between
ground and neutral lines < 400 mV
Compressed air > 400 Kpa (> 50 psi)
Dimensions Mass spectrometer system:
81 cm (width) x 78 cm (depth) x 92 cm (height)
31.9 inch x 30.7 inch x 36.2 inch 
Weight 300 kg (660 lb.), basic system

DELTA V Footprint and Dimensions (in cm)
DELTA V Footprint 1
DELTA V Footprint 2


<= Previous Page => Main DELTA V Page => First Page