TC/EA

  TC/EA



 Since many years on-line isotope ratio analysis of organic bulk samples for 13C/12C, 15N/14N and 34S/32S can be performed rapidly, easily and precisely using conventional Elemental Analyzer -continuous flow IRMS. Measurement of 18O/16O and D/H ratios in organic and inorganic matter had been restricted to off-line sample preparation in the past.

   Product Detail


The new TC/EA (High Temperature Conversion/Elemental Analyzer) is the technological breakthrough bringing the benefits of continuous flow IRMS to on-line oxygen and hydrogen isotope ratio analysis on solid and liquid bulk samples.

  • The TC/EA provides
  • Automated analysis with high sample throughput
  • Small amounts of sample
  • Low blanks and negligible memory
  • Oxygen isotope ratios from organic compounds, water and selected inorganic compounds
  • Hydrogen isotope ratios from organic compounds and water
  • Nitrogen and oxygen isotope ratios from nitrates

Quantitative high temperature conversion, also referred to as pyrolysis, is a new technique in which oxygen present in a compound is converted to CO, and hydrogen contained in a compound is converted to H2.

The process is rapid and quantitative in a reducing environment at high temperature, typically exceeding 1400 °C.

The TC/EA preparation system and ConFlo III interface are outlined in the schematic diagram. This combination facilitates the conversion reaction, separation of reaction gases, transfer into the IRMS and referencing against standard gases. The patented reactor consists of a glassy carbon tube with glassy carbon filling, ensuring that neither sample nor reaction gases can get into contact with oxygen containing surfaces (e.g. Al2O3 ) while at high temperature. Only this technology enables memory-free conversion reactions with no restrictions on compound type.

The reaction gases are separated in an isothermal gas chromatograph, which also is part of the TC/EA. The gases are admitted to the IRMS via the ConFlo III interface. The ConFlo III interface allows automatic sample gas dilution and generation of reference gas pulses, enabling individual referencing of each sample gas peak.

Click here for Trainings information

 
Related Products
 Product# Product Name Image  
 
IQLAAEGAAT
FADEMAIS
 
MAT253
  Select  

Accessories and Related Products
 
 Product# Product Name Image  
 
IQLAAEGAAT
FABHMADZ
 
DELTA V
  Select  
 
Purchase Details
SKU : IQLAAEGAATFAETMALU 
Request Quote
Product Contact
 Sales Contact
  Service Contact