Surveyor 2000

  Surveyor 2000



This product has been discontinued. The recommended replacement is the ASP-2 or Electra GM. Please contact your Sales Rep to discuss alternatives for your particular needs.

For a time-table of spare parts and repair service availability, refer to the Replacement Guide & Support Schedule. For any other questions, please contact our Customer Service department at (800) 274 - 4212.



This product is no longer offered.

   Product Detail


Rugged construction and quality components make it durable and easy to service. Span, HV and calibration pots (one for each range) are clearly marked.

The exclusive HV check assures that the detector is operating at its proper high voltage (a safety feature and critical for operation near the edge of a GM detector plateau). The anti-saturation circuit keeps the meter reading off scale when the detector saturates in a high radiation field, providing an added safety margin.  Automatic dead time compensation assures the accuracy of higher exposure rate readings for linear response on all ranges.

A single 9-volt battery powers the instrument. An optional, license-exempt, uranium check source is available for use with the SURVEYOR 2000.

 
Related Products
 Product# Product Name Image  
 
Surveyor 5
0
 
Surveyor 50
  Select  
 
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Radiation Detected :Alpha, beta, gamma and x-ray, depending upon external GM probe used; gamma and x-ray with internal detector
Detector :Choice of GM probes, external; GM tube, internal
Accuracy :Within 10% of reading for 137 Cs when calibrated according to NRC Reg. Guide 10.8
Energy Response :+ 20% from 40 keV to 1.2 Mev (internal detector)
High Voltage :Electronically stabilized, factory set at 900 V
HV Test :Exclusive self-test to verify detector HV power supply
Connector :MHV
Warm-up Time :None
Saturation :Typically > 1000 R/h on all ranges (with exclusive antisaturation circuit) for most GM probes; > 5 R/h for pancake GM probes