Welcome Guest from United States
Sign In Change Country
  0 Items
Search:
Product Browse Products >>... >> Air Quality Monitoring Solutions >> Source Gas Analyzers
SO2 Gas Analyzer, Model 43i

  SO2 Gas Analyzer, Model 43i



A change is in the air 
The industry?s new best-of-breed.  Our customers told us exactly what they?re looking for in a gas monitoring solution: reliability, simplicity, ease of use.  The new iSeries platform delivers on all counts - and then goes a step farther.
A core product in Thermo?s new iSeries product line is the Model 43i SO2 analyzer.  

   Product Detail


Using pulsed fluorescence technology, the Model 43i measures the amount of sulfur dioxide in the air up to 10ppm.  The pulsing of the U.V. source lamp serves to increase the optical intensity whereby a greater U.V. energy throughput and lower detectable SO2 concentration are realized.  Reflective bandpass filters, as compared to commonly used transmission filters, are less subject to photochemical degradation and more selective in wavelength isolation.  This results in both increased detection specificity and long term stability.

This state-of-the-art gas analyzer offers features such as an ethernet port as well as flash memory for increased data storage.  Ethernet connectivity provides efficient remote access, allowing the user to download measurement information directly from the instrument without having to be on-site.

You can easily program soft-keys to allow you to jump directly to frequently accessed functions, menus or screens  The larger interface screen can display up to five lines of measurement information while primary screen remains visible.

Key Features

  • Ethernet connectivity for efficient remote access
  • Enhanced user interface with one button programming and large display screen
  • Flash memory for increased data storage and user downloadable software
  • Enhanced electronics design optimizes product commonality
  • Improved layout for easier accessibility to components

Comprehensive Service Solutions

To maintain optimal product performance, you need immediate access to experts worldwide, as well as priority status when your air quality equipment needs repair or replacement.  Thermo Scientific offers comprehensive, flexible support solutions for all phases of the product lifecycle.  Through predictable, fixed-cost pricing, Thermo services help protect the return on investment and total cost of ownership of your Thermo Scientific air quality products.

 
Related Products
 Product# Product Name Image  
  43C   Model 43C SO2 Analyzer - See Model 43i
  Select  
 
Purchase Details
SKU : 43i 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Firmware Downloads for Air Quality Instruments
  • Software Downloads for Air Quality Instruments
  • Specifications:
    Preset Ranges :0-0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 ppm; 0-0.2, 0.5, 1, 2, 5, 10, 20, and 25 mg/m3
    Extended Ranges :0-0.05, 1, 2, 5, 10, 20, 50 and 100 ppm; 0-2, 5, 10, 20, 50, 100, 200, and 250 mg/m3
    Custom Ranges :0-0.05 to 10 ppm; 0-0.2 to 250 mg/m3
    Zero Noise :1.0 ppb RMS (10 second averaging time), 0.5 ppb RMS (60 second averaging time), 0.25 ppb RMS (300 second averaging time)
    Lower Detectable Limit :2.0 ppb (10 second averaging time), 1.0 ppb (60 second averaging time), 0.5 ppb (300 second averaging time)
    Zero Drift (24 hour) :Less than 1 ppb
    Span Drift (24 hour) :+/-1%
    Reponse Time :80 seconds (10 second average time); 110 seconds (60 second average time); 320 seconds (300 second average time)
    Precision :1% of reading or 1 ppb (whichever is greater)
    Linearity :+/-1% full scale < 100ppm
    Sample Flow Rate :0.5 liters/min. (standard); 1 liter/min. (optional)
    Interferences (EPA levels) :< lower detectable limit except for the following: NO < 3 ppb, M-Xylene < 2 ppb, H2O < 2% of reading
    Operating Temperature :20°C - 30°C
    Power Requirements :100 vac, 115 vac, 220-240 vac +/-10% @ 165W
    Size and Weight :16.75?(W) x 8.62?(H) x 23?(D), 48 lbs. (22.2 kg)
    Outputs :Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail Indication (standard). 0-20 or 4-20 mA Isolated Current Outout (optional)
    Inputs :16 Digital Inputs (standard), 8 0-10vdc Analog Inputs (optional)