Orion 5-Star Plus Benchtop Multimeter


  • Ability to turn off BOD stirrer
  • Solution and membrane temperature printed with measurement
  • Solution and membrane temperature displayed
  • Ability to calibrate both solution and membrane temperature
  • DO slope now on printout
  • Third decimal point for cell constant entry
  • Added 5 and 10°C as reference temperature options
  • Added second decimal point to temperature compensation
  • Ability to calibrate conductivity temperature sensor
  • Now with ability to operate an auto-sampler
   Product Detail


This highly versatile, competitively-priced, multi-channel, multi-display meter is easy to set up and use while delivering exceptional, long-life performance for demanding R&D and QA/QC applications. The rugged meter offers a full complement of GLP-related features including 21 CFR Part 11 software.

Optional 1 to 5 point calibration for pH, direct ion and conductivity measurement ensures a high level of accuracy even at the extremes of concentration; optional calibration to Winkler Titration for BOD measurements.

SMART STABILITYTMand SMART AVERAGINGTM functions optimize the measurement process. You can record and store up to 10 setup parameters and calibrations and download additional methods and software updates via the Web.

Additional features:

  • Improved stability algorithm to increase calibration and measurement accuracy
  • Average slope and slope of segment printed with measurement
  • Ability for ATC temperature calibration
  • Increased data log points: 500
  • Ability to turn off data log
  • Ability to print or view last 10 calibrations
  • New Individual method passwords
  • 21 CFR Part 11 Compliant
  • Bright LCD display with ready indication
  • Convenient built-in Quick-Start Guide
  • pH, ORP, DO, ISE and conductivity measurements in one compact unit
  • Stores up to 10 methods with password protection
  • IP54 dustproof and splashproof
  • BNC pH, ORP and ISE connectors and mini DIN locking DO and conductivity probe connectors
  • Automatic or manual pH calibration with 1 to 5 calibration points
  • Auto-buffer-recognition of NIST/US and DIN buffers
  • Uses polarographic DO probe
  • Automatic or manual temperature compensation
  • Manual ISE calibration with 1 to 5 calibration points
  • Point-to-point ISE calibration, high and low concentration range
  • Automatic blank correction
  • Automatic barometric pressure compensation
  • Auto-read/hold function; GLP IQ/OQ documentation included
  • Advanced self-tests on meter operation, self-diagnostics
  • Optional USB interface
  • CE, CSA, UL, TÜV and FCC Class A limits approved
  • Universal power adapter for worldwide operation
  • Three-year replacement warranty.
 Orion 5-Star Plus Benchtop Multimeter
 Product# Price  Description  
       
 1119000      -    Meter Only   SELECT  
 1119001      -    Full Benchtop Multimeter Kit with probes and buffers   SELECT  
 1010152      -    Water Conductivity 645 Meter Application Kit   SELECT  
 1119201      -    Meter kit with stand and probes   SELECT  
 1010113      -    Wastewater Sulfide Meter Application Kit   SELECT  
 
Related Products
pH Electrode Cleaning and Storage Solutions
 Product#  
   
 910060  
SELECT   
pH Buffer Bottles
 Product#  
   
 910460  
SELECT   
pH Buffer perpHect® Individual Pouches
 Product#  
   
 910410  
SELECT   
 911010  
SELECT   
 910710  
SELECT   
Conductivity Standards, Solutions and Accessories
 Product#  
   
 011007  
SELECT   
DO Maintenance Kits, Membranes, Solutions and Accessories
 Product#  
   
 080513  
SELECT   
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
pH :
Range/ Resolution :–2.000 to 19.999 or 0 to 14 in aqueous solutions / 0.1,0.01,0.001
Relative accuracy :+/- 0.002
SMART STABILITY™ and SMART AVERAGING™ :Resolution dependent / automatic
AUTO-CAL™ and/or Manual Calibration :1 to 5 points; US/NIST, DIN, and custom buffers recognized
ISE :
Range/Resolution :0 to 19900 / 1,2, or 3 significant digits
Relative Accuracy :+/- 0.2mV or 0.05%, whichever is greater
SMART STABILITY™ and SMART AVERAGING™ :Resolution dependent / automatic
Units :M, mg/L, %, ppb, or no units
# ISE Cal points :1 to 5
Special ISE Calibration Features :Linear point to point, selectable auto-blank non-linear, low concentration range
mV :
Range – mV / RmV :± 1999.9
ORP :EH autocalibration to standard hydrogen electrode
Resolution :0.1
Relative Accuracy :± 0.2 mV or 0.05%, whichever is greater
Conductivity :
Range :0.000 to 3000 mS/cm, auto resolution with cell constant dependence
Resolution :4 significant digits down to 0.001 µS/cm, cell constant dependent
Relative Accuracy :0.5% ± 1 digit or 0.01 µS/cm, whichever is greater
Cell Constant :0.001 to 199.9
Cell Compatibility :2 Cell, 4 Cell, Planar Cell
SMART STABILITY™ and SMART AVERAGING™ :Resolution dependent / automatic
DirectCal™ / AUTO-CAL™ :1 to 5 points / 1 to 3 points
Resistivity :
Range / Resolution :0.0001 to 100 Megohm / Auto resolution
Relative Accuracy :0.5 % ± 1 digit
Salinity :
Range / Resolution :0.1 to 80.0 ppt NaCl equivalent, 0.1 to 42ppt practical salinity / 0.1
Relative Accuracy :± 0.1 ± 1 digit
TDS :
Range / Resolution :0 to 19999 / 1 mg/L
Relative Accuracy :± 0.5 % ± 1 digit
Adjustable TDS Factor :0.00 to 10.00
DO :
Range / Resolution Concentration :0.00 to 90.0 mg/L / 0.0/0.00
Range / Resolution % Saturation :0.0 to 600% / 0/0.0
Relative Accuracy Concentration / % Saturation :± 0.5% ±1 digit
SMART STABILITY™ and SMART AVERAGING™ :Resolution dependent / automatic
Auto Calibration or Manual Input :Water Saturated Air and Air Saturated Water, Winkler
Calibration to Winkler Titration for BOD :Available as needed
Salinity Factor :0 to 45 ppt
Barometric Pressure Range :450 to 850 mm Hg
Probe Type :Polarographic
Temperature :
Range / Resolution :-5 to 105ºC / 0.1 up to 99.9ºC, 1.0 over 99.9ºC
Relative Accuracy :± 0.1 ºC
Temperature Compensation :DO (auto); Cond (nLF, linear 0.0 to 10.0% /ºC); pH (auto/manual)
Display :Custom backlit LCD, for easy viewing
Inputs/Outputs :
Sensor Inputs :Conductivity-miniDIN - Locking; DO-miniDIN - Locking; ATC-miniDIN-Locking; pH/ISE - BNC; Reference-Pin tip; Stirrer-Phono-jack
# Of Data Logging Points :200 points by time, stability or manual read with time/date stamp
Calibration Log :Last 10 Calibrations
User Method Storage :10 Saved
Advanced Self-test and Diagnostics :YES
GLP Electrode Verification :YES
Password Protection :YES
Internet Software Updates / Downloadable Methods :YES
21 CFR Part 11 Software Package :YES
Power :Universal power adapter and battery power 4 x AA batteries
Regulatory and Safety :CE, CSA, UL, TÜV, FCC Class A limits
Environmental Operating Conditions :
Ambient Operating Temperature :5 to 45ºC
Relative Humidity :5 to 85% Non-Condensing
IP Rating :IP54, splashproof and dustproof
Warranty :36 months (from date of purchase)