Partisol Model 2000

  Partisol Model 2000



 The Partisol Model 2000 Air Sampler from is a flexible manual air sampling platform for the measurement of particulate matter and organics in ambient air.
   Product Detail


The Partisol Model 2000 Air Sampler is a flexible manual air sampling platform for the measurement of particulate matter and organics in ambient air. The hardware is a USEPA method for the sampling of PM-10 (RFPS-0694-098), and can be easily fitted with sample inlets for PM-2.5, PM-1 and TSP monitoring. It has also been tested by the TUV in Essen, Germany, for compliance with the European EN12341 PM-10 sampling method. The sampler represents a major advance in air sampling technology, incorporating such innovations as a hub and satellite configuration, flexible microprocessor control, and ActiVol flow control.

Major air pollution monitoring agencies, private enterprises and consulting companies worldwide employ the Partisol Sampler to generate quality-assured particulate mass measurements of ambient air. It can be configured in a variety of ways to respond to specialized air monitoring needs with minimal expense and effort. Its unique feature set has prompted an increasing number of organizations to supplant high volume samplers with Partisol.

  • Choice of standard 47 mm diameter U.S. EPA-approved filter media for regulatory PM-10 measurements.
  • Exposed collection filters can be analyzed for using standard laboratory techniques such as AA and ICP.
  • Filter exchange made easy by quick-exchange filter mechanism and filter cassettes.
  • ActiVol flow control system maintains a constant user-selectable volumetric flow rate by incorporating a mass flow controller, ambient temperature sensor and ambient pressure sensor. The system reports sampled volumes in standard units (standard m3).
  • Collection hardware includes single 47 mm filters, polyurethane foam (PUF) samplers and multi-stage filter packs.
  • Direct flow audit and calibration procedures using NIST-traceable standards.
  • Hub and satellite configuration allows up to three cost-effective satellite units to be controlled by a Partisol hub sampler.
  • Microprocessor allows for simple setup of 24-hour and conditional. Operating data are stored internally for viewing and downloading.
  • RS232 and analog interfaces provide advanced communication capabilities.
  • Low noise--appropriate for indoor monitoring.
  • Low maintenance requirements through the use of durable components and long-life vacuum pump.
 
Related Products
 Product#  Product Name Image  
 FTS   Streamline FTS 
  Select  
 Streamline Pro   Streamline Pro 
  Select  

Accessories and Related Products
 
 Product#  Product Name Image  
 Dichotomous Partisol-Plus 2025-D   Dichotomous Partisol-Plus 2025-D 
  Select  
 Partisol FRM, Model 2000   Partisol FRM, Model 2000 
  Select  
 Partisol-Plus, Model 2025   Partisol-Plus, Model 2025 
  Select  
 
Purchase Details
SKU : Partisol Model 2000 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Sample Flow :5 to 18 l/min
Filters :47mm Quartz, Teflon-coated glass fiber or Teflon membrane
Interfaces :RS-232 interface, 2 analog inputs, windvane/anemometer connector, 1 analog output
Particle Size Cutpoints :PM-10, PM-2.5, PM-1 or TSP
Operating Range :-40 to 50 °C. Conditions below -25 °C require optional insulating jacket
Power Requirements :2.2A@120VAC, 1.1A@240VAC
Hub Unit Dimensions :w: 16" (41 cm) x D: 13" (33 cm) x H: 24" (61 cm). Inlet adds 31" (79 cm) to height
Hub Unit Weight :70 lb (32 kg)
Satellite Unit Dimensions :W: 10" (25 cm) x D: 13" (33 cm) x H: 15" (38 cm). Inlet adds 31" (79 cm) to height