Welcome Guest from United States
Sign In Change Country
  0 Items
Search:
Product Browse Products >>... >> Air Quality Monitoring Solutions >> Source Gas Analyzers
Mercury Analyzer, Model 80i

  Mercury Analyzer, Model 80i



It's time for a new standard in mercury emissions monitoring.
Part of the Mercury Freedom System, the Model 80i is one of the four major components of this state-of-the-art system.

   Product Detail


The Model 80i utilizes an advanced cold vapor atomic fluorescence technology. This technology enables the Model 80i to provide continuous sample measurement, with no additional gases or accumulation required - and virtually no interference from SO2. High sensitivity allows high sample dilution reducing moisture, heat, and interfering pollutants.

Designed to meet the provisions of 40CFR Parts 60 and 75, the Mercury Freedom System provides a complete mercury monitoring solution that reports elemental, ionic, and total mercury in exhaust stacks from both coal-fired boilers and waste incinerators.

The Model 80i also offers the benefits of the new iSeries gas analyzer platform.

The iSeries platform features flexible communications, increased serviceability, and an easier to use interface. Ethernet connectivity provides efficient remote access, allowing the user to download measurement information directly from the instrument without having to be on site. Captive hardware and slide-apart modules allows easy access for service and periodic maintenance.

You can easily program soft-keys to allow you to jump directly to frequently accessed functions, menus or screens. The larger interface screen can display up to five lines of measurement information while the primary reporting screen remains visible.

Key Features:

  • Proprietary cold vapor atomic fluorescence method
  • Reports elemental, ionic, and total mercury
  • True, real-time monitoring means no batch processing
  • Direct measurement eliminates carrier gas
  • High sensitivity allows greater dilution ratiosMethod reduces interferences
  • Fast intuitive navigation

 
Related Products
 Product# Product Name Image  
 
Mercury Fr
eedom
 
The Mercury Freedom System
  Select  
 
Purchase Details
SKU : 80i 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Firmware Downloads for Air Quality Instruments
  • Specifications:
    Elemental Hg Measurement Range :0-50µg/m³ (effective range before dilution)
    Zero Noise :1ng/m³ (300 second average)
    Lower Detectable Limit :2ng/m³ (300 second average)
    Zero Drift (24 hour) :3ng/m³
    Response Time :90 seconds (60 second average time)
    Linearity :+/- 1% full scale
    Sample Flow Rate :0.25 slpm per channel
    Interference Rejection (SO2) :8.4 x 10 to the 8th power:1 SO2 interference below LDL
    Interference Rejection (NOx) :4.2 x 10 to the 9th power: 1 NO interference below LDL
    Operating Temperature :5 - 40°C
    Power Requirements :100 vac, 115 vac, 220-240 vac +/-10% @ 275W
    Size and Weight :16.75"(W) x 8.62"(H) x 23"(D), 49 lbs. or 425mm (W) x 219mm (H) x 584mm (D), 22.2 kg
    Outputs :Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays and Power Failure Indication. 0-20 or 4-20 mA Isolated Current Output
    Inputs :16 Digital Inputs, 8 0-10vdc Analog Inputs