The Thermo Scientific Orion model 2111XP Sodium Analyzer offers unmatched reliability in analyzing critical sample streams throughout the power/steam generation and industrial water industry. It is ideally suited to meet the demanding needs of high purity water measurements and high acid-cation exchange applications all in one system all from one of the most trusted name in sodium monitoring.
Product Detail
Provides early warning detection of sodium
Simple step-by-step scrolling instruction
Measurements at a glance with visible backlit display
NEW Orion sodium electrode technology limits calibration cycles
Minimize operator time and maintenance without use of complicated moving parts or pumps
Detection limit of 0.1 ppb the Orion 2111XP offers flexible application packages ideally suited for continuous sodium monitoring in a wide variety of samples.
Cat. No.
Reagent Packages
Description
211190FOR*
Reagent-less Diisopropylamine Reagent
(1) x 0.8 L empty bottle for use with DIPA reagent,, complete with (1) diffusion tube assembly (211194)
211191FOR*
Reagent-less Diisopropylamine Reagent
(6) x 0.8 L empty bottles for use with DIPA reagent, complete with (6) diffusion tube assembly (211194)
For freshly etched electrode, during calibration: Reached 95% of final reading within 2 minutes of injecting a standard solution
Units displayed :
ppb, ppm auto ranging
Display :
Custom backlit LCD with temperature, concentration, mV, error codes, menu driven prompts
Power :
85-132 or 170-264 V AC
Regulatory and Safety :
CE, CSA, UL, FCC Class A limits
Case Material :
Valox 364
Temperature :
5 to 45?C
Total acidity :
less than 250 ppm CaCO3
Inlet pressure :
8 to 100 psig
Flow Rate :
40 ml/min nominal set by pressure regulator (DIPA and Ammonia applications); 20 ml/min nominal set by pressure regulator (cation/high acid applications)
Sample Inlet :
1/4" NPTF tube fitting
Sample Drain :
3/4" NPT male
Grab Sample Supported :
Yes
Analog Output :
Galvanically Isolated
# of Analog Outputs :
2, one dedicated to sodium, one to temperature; shared GND
Output selections :
0 -20 mA or 4 -20 mA
Programmable range :
Yes
Log & Linear output options :
Yes, user selectable
Alarm outputs :
3
# of relay outputs :
3
Max relay load :
250 Vac, 5 A, 30 Vdc
Min./Max. Value Alarm :
Yes
Error & Offline/Cal Alarms :
Yes
Programmable min & max values :
Yes
Electronics Dimensions :
144 x 144 x 186 mm
Entire System Dimensions :
65 x 45 x 27 cm (26" x 17" x 11"). Fits on 1811EL/1811AO panel mounting footprint