CyberScan PC 510


The multi-parameter CyberScan PC 510 microprocessor-based bench meter incorporates many advanced features to meet today's requirements. Push-button calibration ensures simple operation while configurations can be customized through its SETUP function. Meter comes with epoxy-body pH double junction electrode and conductivity electrode with a built-in temperature sensor, complete with an integral electrode arm.

*This product is not sold in North America

   Product Detail


  • Multi-parameter meters measures pH, Conductivity and TDS in single meter
  • User-Customization for more accurate measurements with programmable functions like stability, electrode diagnosis - pH slope/offset/mV base value, etc.
  • Multi-Point Push-Button Calibration (up to 5 points for pH/Conductivity/TDS) for unprecedented full-range accuracy
  • Selectable USA or NIST pH Buffer Standards ensures flexibility in calibration
  • Auto-Buffer Recognition eliminates guesswork and mistakes during calibration
  • Automatic or Manual Temperature Compensation
  • Hold Function freezes reading momentarily for easy viewing and recording
  • Large Easy-to-Read Custom LCD Display for optimum viewing of dual measurements of pH, Conductivity and TDS with corresponding temperature (oC/oF)
  • Stability Indicator prompts whenever reading stabilizes
  • Easy Interchangeability of probes that can be replaced with other specialized electrodes for specific application
  • Slide-out Instruction Card for quick reference
  • Electrode With Built-In ATC designed for minimal air bubble entrapment during measurement
  • Detachable Electrode Stand organizes multiple electrodes neatly and holds them firmly in place
 CyberScan PC 510
 Product# Price  Item  
       
 EC-PC510/13S      -    CyberScan PC 510 Bench pH/Conductivity/TDS Meter with pH electrode (EC-FC72522-01B), Conductivity electrode   SELECT  
 EC-FC72522-01B      -    General Purpose Plastic-body Double Junction Gel-Filled pH Combination Electrode, 12x110mm, BNC connector,   SELECT  
 ECCONSEN91W      -    2 Stainless Steel Rings Ultem-body Electrode with ATC, cell constant K=1.0, 14x110 mm, 6-pin connector, 1m cable length   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Conductivity :0 to 20.00, 200.0, 2000 µS/cm; 20.00, 200 mS/cm
TDS Range :0 to 10.00, 100.0, 1000 ppm; 10.00, Max. 100 ppt (depending on factor setting)
Resolution :0.05% of Full Scale
Accuracy :± 1% Full Scale + 1 digit
Temperature Range :-10.0 to 110.0 °C (14.0 to 230 °F)
Resolution & Accuracy :0.1 °C & ± 0.5 °C (0.9 °F)
Calibration for Conductivity :Automatic (single or multi-point up to 4 points) / Manual (up to 5 points)
Calibration for TDS :Manual (up to 5 points - 1 per range)
Normalization Temperature :15.0 to 30.0 °C
Cond. to TDS Conversion Factor :0.40 to 1.00 (selectable)
Cell Constant :0.1, 1.0 or 10.0 (selectable)
Temperature Coefficient :0.0 to 10.0 %/°C (adjustable)
Memory :50 Data Sets
Special Functions :Auto-ranging; Auto-calibration, Auto-hold; Stability Indicator & Self-diagnostic
Power :9VDC, using AC adapters 110VAC or 220VAC, 50/60 Hz