alpha-pH2000 1/2 DIN Wall Mount/Panel Mount pH/ORP Analyzer

  alpha-pH2000 1/2 DIN Wall Mount/Panel Mount pH/ORP Analyzer



In the ORP mode, the alpha pH 2000 is able to measure in mV or as % concentration, with independent calibration modes.  All parameters can be operated in the asymmetrical or symmetrical mode.

With a pH or ORP controller side-by-side, controllers can be operated in a master-slave operation where pH and ORP must be measured and controlled simultaneously, e.g. electroplating or swimming pool applications.  Under this setup, pH is always maintained and controlled as a first priority.  Only when pH readings are within set limits, will the ORP controller be allowed to operate the pumps. 

pH calibration is quick and easy with push buttons.  Calibration buffer sets include US, NIST and DIN standards.  Electrode condition is updated and displayed after each calibration.

*This product is not sold in North America


   Product Detail


  • Built-In Programmable for On/Off, Proportional, or PI (Proportional integral control) - ideal for most demanding precision process control applications
  • User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement
  • Automatic Calibration with Auto-Buffer Recognition eliminates mistakes during calibration
  • Symmetrical Mode Operation eliminates electronic noise problems when used with solution ground
  • Backlit Display offers reading in low-light conditions
  • Universal Power Supply Input from 85 to 250 VAC/DC allows use at multiple sites and withstands voltage fluctuations
  • Rugged NEMA 4X housing protects controller from the elements and dirty industrial applications
  • Two Level Password Protection prevents unauthorized tampering with settings
  • 0 to 2000 Second Time Delay Adjustment on control and alarm delays
  • Hold Relay for use with float switches/flow switches and other controllers as a failsafe function
  • Two Galvanically Isolated Scaleable 0-20/4-20 mA Outputs for pH/ORP and temperature
  • Wash Contact Relay controls electrodes cleaning systems at desired duration and frequency
  • Choice of Glass or Antimony Electrode for general purpose or hydrofluoric acid applications
  • Adjustable Hysteresis (Dead Band) prevents rapid contact switching near set point
  • Non-Volatile Memory retains all stored parameters and calibration data even if power fails
  • Large Dual Display shows pH (or ORP) with temperature simultaneously - features clear multiple icons, set points, and status messages
  • Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection
  • Easy Installation and Wiring with detachable plug-in connectors
  • Panel-mounted version is also available

Applications

General: Useful for any batch or on-line type application that requires accurate pH or ORP control.

Water Purification/Treatment: Use for batch and on-line control of incoming process water, rinse water treatment, recirculating system and waste water treatment.

Industrial: Ideal for chemical processing, food processing. aquarium, pharmaceutical, hydroponics and waste control industries.

Regulatory: Hook to recorder to document data for regulatory compliance

 alpha-pH2000 1/2 DIN Wall Mount/Panel Mount pH/ORP Analyzer
 Product# Price  Item  
       
 TSPHCTP2000W      -    alpha pH / ORP Controller / Transmitter (Wall-mount)   SELECT  
 TSPHCTP2000P      -    alpha pH / ORP Controller / Transmitter (Panel-mount)   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
pH Range :-2.00 to 16.00 pH
Resolution & Accuracy :0.01 pH & ±0.01 pH
mV Range :-1000 to 1000 mV / 0 to 100% (for -100 to 100 mV)
Resolution & Accuracy :1 mV / 0.1% & ±1 mV / ±0.2%
Temperature Range :-10.0 to 110 °C (14.0 to 230.0 °F)
Resolution & Relative Accuracy :0.1 °C / °F & ±0.5 °C / ±1.0 °F
Temperature Sensor :Pt100/ Pt1000 (jumper selectable); 2 or 3 wire
Temperature Compensation :Auto / Manual (Reference at 25°C)
Integral Time :0 to 999.9 minutes
Switching pH Hysteresis :0.1 to 1 pH
Function (Switchable) :P/PI control (Pulse Length / Pulse Frequency); Limit Control; Off
Adjustable Period with Pulse Length Controller :0.5 to 20 seconds
Adjustable Period with Pulse Frequency Controller :60 to 120 pulses/min
Pickup / Dropout Delay :0 to 2000 seconds
Wash Cycle :0.1 to 200.0 hours
Wash Duration :1 to 2000 seconds
Contact Outputs, Controller :1 SPDT, 3 SPST relays
Switching Voltage / Current / Power :Max 250 VAC / Max 3A / Max 600VA
Function (Switchable) :Latching / pulse
Pickup Delay :0 to 2000 seconds
LCD :UV coat, backlit 14 segments display with symbols for status information
Backlight :On/Off selectable with four levels of brightness control
Emitted Interference / Immunity to Interference :According to EN 61326
Ambient Temperature Operating Range :0 to 40 °C
Maximum Relative Humidity :80% up to 31 °C decreasing linearly to 50% at 40 °C
Input :80 to 250 VAC/DC 50/60 Hz Approx. 10 VA
Main Fuse :250 mA, Anti-surge, S504 BUSSMAN
Pollution Degree :2
Transient Overvoltage Category :II
Transmitter Function :Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated
CU 22 Function :22 mA current output
Isolated Output Voltage :+12V ±0.5 V (Max 50 mA)
Hold Function Switch :To freeze output current and deactivate control relays
Load :Max 600 W
pH Input :BNC (1012 impedance); Asymmetrical/ Symmetrical
Connection Terminal (Wall Mount) :3-pin, 8-pin, 9-pin and 13-pin terminal, detachable blocks
Connection Terminal (Panel Mount) :3-pin, 9-pin and 19-pin terminal, detachable blocks
Dimensions (W x H x D) :

Wall-mount 144 x 144 x 110 mm

Panel-mount 96 X 96 X 175 mm

Weight :

Wall-mount 745g (Unit) / 1100g (Packed)

Panel-mount 950g (Unit) / 1100g (Packed)

Environmental Rating :

Wall-mount IP66 (NEMA 4X)

Panel-mount IP54 (front panel)