alpha-CON2000 Wall Mount/Panel Mount pH/ORP Analyzer

  alpha-CON2000 Wall Mount/Panel Mount pH/ORP Analyzer



Seven ranges in a single controller allows the controller to be used in a wide variety of applications.  Maximum measurement range is 1000 mS/cm.  Accuracy of ±1 is achievable for most ranges, while ±2% for measurements above 500 mS/cm.  

New 4-cell electrode system eliminates polarizing effects ensuring more accurate measurements in highly conductive samples.

*This product is not sold in North America


   Product Detail


  • Seven ranges in a single controller allows the controller to be used in a wide variety of applications with the maximum range at 1000 mS/cm
  • User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement
  • Choice of 4-cell or 2-cell electrode system
  • Programmable for On/Off, Proportional, or PI (Proportional Integral Control) - ideal for most demanding precision process control applications
  • Backlit Display offers reading in low-light conditions
  • Universal Power Supply Input from 85 to 250 VAC/DC allows use at multiple sites and withstands voltage fluctuations
  • Two Galvanically Isolated Scaleable 0-20/4-20 mA Outputs for Conductivity and temperature
  • Wash Contact Relay controls electrodes cleaning systems at desired duration and frequency
  • Adjustable Hysteresis (Dead Band) prevents rapid contact switching near set point
  • Non-Volatile Memory retains all stored parameters and calibration data even if power fails
  • Large Dual Display shows measurement with temperature simultaneously - features clear multiple icons, set points, and status messages
  • Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection
  • Easy Installation and Wiring with detachable plug-in connectors
  • Panel-mounted version is also available

Applications

General: Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator.

Industrial: Use in applications involving agriculture, chemical processing, boiler and water heaters, wafer-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries, and bleach manufacturing.

Water Purification/Treatment: Use to treat batches of incoming process water, ultrapure water, boiler and feed water control.

Regulatory: Hook to recorder to document data for regulatory compliance.

 alpha-CON2000 Wall Mount/Panel Mount pH/ORP Analyzer
 Product# Price  Item  
       
 TSCONCTP2000W      -    alpha CON 2000 Conductivity Controller / Transmitter (Wall-mount)   SELECT  
 TSCONCTP2000P      -    alpha CON 2000 Conductivity Controller / Transmitter (Panel-mount)   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Conductivity Ranges/ Resolution :

.. to 2.000 µS/cm 0.001 µS/cm

.. to 20.00 µS/cm 0.001 µS/cm

.. to 200.0 µS/cm 0.1 µS/cm

.. to 2000 µS/cm 1 µS/cm

.. to 20.00 mS/cm 0.01 mS/cm

.. to 200.0 mS/cm 1 mS/cm

Accuracy :±1 % of full scale reading (±2 % > 500 mS/cm)
Temperature Range :-10.0 to 125 °C (14.0 to 257.0 °F)
Resolution & Relative Accuracy :0.1 °C / °F & ±0.5 °C / ±1.0 °F
Temperature Sensor :Pt100/ Pt1000 (jumper selectable); 2 or 3 wire
Temperature Compensation :Auto / Manual (Reference at 25°C)
Integral Time :0 to 999.9 minutes
Switching Hysteresis :0 to 10 % of full scale
Function (Switchable) :P/PI control (Pulse Length / Pulse Frequency); Limit Control; Off
Function (Switchable) :P/PI control (Pulse Length / Pulse Frequency); Limit Control; Off
Adjustable Period with Pulse Length Controller :0.5 to 20 seconds
Adjustable Period with Pulse Frequency Controller :60 to 120 pulses/min
Pickup / Dropout Delay :0 to 2000 seconds
Wash Cycle :0.1 to 200.0 hours
Wash Duration :1 to 2000 seconds
Contact Outputs, Controller :1 SPDT, 3 SPST relays
Switching Voltage / Current / Power :Max 250 VAC / Max 3A / Max 600VA
Function (Switchable) :Latching / pulse
Pickup Delay :0 to 2000 seconds
LCD :UV coat, backlit 14 segments display with symbols for status information
Backlight :On/Off selectable with four levels of brightness control
Emitted Interference :According to EN 61326
Immunity to Interference :According to EN 61326
Ambient Temperature Operating Range :0 to 40 °C
Maximum Relative Humidity :80% up to 31 °C decreasing linearly to 50% at 40 °C
Input :80 to 250 VAC/DC 50/60 Hz Approx. 10 VA
Main Fuse :315 mA, Anti-surge, S504 BUSSMAN
Pollution Degree :2
Transient Overvoltage Category :II
Transmitter Function :Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated
CU 22 Function :22 mA current output
Isolated Output Voltage :+12V ±0.5 V (Max 50 mA)
Hold Function Switch :To freeze output current and deactivate control relays
Load :Max 600 W
Conductivity Input :Screw terminal
Connection Terminal (Wall Mount) :3-pin, 8-pin, 9-pin and 13-pin terminal, detachable blocks
Connection Terminal (Panel Mount) :3-pin, 9-pin and 19-pin terminal, detachable blocks
Dimensions (W x H x D) :

Wall-mount: 144 x 144 x 110 mm

Panel-mount: 96 X 96 X 175 mm

Weight :

Wall-mount: 745g (Unit) / 1100g (Packed)

Panel-mount: 950g (Unit) / 1100g (Packed)

Environmental Rating :

Wall-mount: IP66 (NEMA 4X)

Panel-mount: IP54 (front panel)