alpha DO2000 1/2 DIN Series Dissolved Oxygen Analyzer/Controller

  alpha DO2000 1/2 DIN Series Dissolved Oxygen Analyzer/Controller



Incorporating the superior Galvanic oxygen probe, operating the alpha DO 2000 controller is immediate.  With the Galvanc probe, there is no start-up or 'warm-up' time.  The probe only takes 40 to 50 seconds to attain 95% of actual readings.  Calibration is quick and easy using the atmospheric air as the calibration media for 100% air saturation.

*This product is not sold in North America


   Product Detail


  • Available in both Galvanic (αlpha DO2000) and Polarographic (αlpha DO2000PG) versions
  • Programmable for On/Off, Proportional, or PI (Proportional integral CONTROL) - ideal for most demanding precision process control applications
  • Calibration is quick and easy using the atmospheric air as the calibration media for 100% air saturation
  • User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement
  • Includes Barometer pressure, Salinity, Temperature Compensation
  • 0 to 2000 Second Time Delay Adjustment on control and alarm delays
  • Non-Volatile Memory retains all stored parameters and calibration data even if power fails
  • Backlit 14 segments display with symbols for status information and On/Off selectable with four levels of brightness control
  • Adjustable Hysteresis (Dead Band) prevents rapid contact switching near set point
  • Three-Level Password Protection prevents unauthorized tampering with settings
  • Wash Contact Relay controls electrodes cleaning systems at desired duration and frequency
  • Universal Power Supply Input from 85 to 250 VAC/DC allows use at multiple sites and withstands voltage fluctuations
  • Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection
  • Dual Display shows Dissolved Oxygen value with temperature simultaneously - features clear multiple icons, set points, and status messages
  • Rugged NEMA 4X housing protects controller from the elements and dirty industrial applications
  • Easy Installation and Wiring with detachable plug-in connectors

Applications

General: Useful for monitoring temperature and control of DO in wastewater treatment and aquaculture applications. Use with valves, blowers or aerators to ensure proper system operation.

 alpha DO2000 1/2 DIN Series Dissolved Oxygen Analyzer/Controller
 Product# Price  Item  
       
 TSDOCTP2000W      -    alpha DO 2000 Controller / Transmitter (Galvanic; wall mount)   SELECT  
 TSDOCTP2000WPG      -    alpha DO 2000 Controller / Transmitter (Amperometric, wall mount)   SELECT  
 TSDOCTP2000PPG      -    alpha DO 2000 Controller / Transmitter (Amperometric, panel mount)   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
DO Range :Galvanic: 0.00 – 25.00 mg/l or 0.00 – 300% Saturation Amperometric: 0.00 – 20.00 mg/l or 0.00 – 200% Saturation
Resolution :0.01 mg/l or ppm or 0.1 %
Accuracy :Galvanic: ±1.5 % full scale Amperometric: ± 1% full scale
Temperature Range :-10.0 to 125 °C (14.0 to 257.0 °F)
Resolution & Relative Accuracy :0.1 °C / °F & ±0.5 °C / ±1.0 °F
Temperature Sensor :Galvanic: Pt100/ Pt1000 (jumper selectable); 2 or 3 wire Amperometric: NTC 22 kΩ Thermistor
Temperature Compensation :Auto / Manual
Salinity Compensation :0.0 to 50.00 ppt (Manual setting and Automatic correction)
Pressure Compensation :0.740 to 3.000 Bar; 555 to 2250 mmHg; 10.73 to 43.51 PSI (Manual setting and Automatic correction)
Integral Time :0 to 999.9 minutes
Switching Hysteresis :Galvanic: 0 to 10% full scale Amperometric: 0.1 to 1.0 mg/l or 1.0 to 10.00 %
Function (Switchable) :P/PI control (Pulse Length / Pulse Frequency); Limit Control; Off
Adjustable Period with Pulse Length Controller :0.5 to 20 seconds
Adjustable Period with Pulse Frequency Controller :60 to 120 pulses/min
Pickup / Dropout Delay :0 to 2000 seconds
Wash Cycle :0.1 to 200.0 hours
Wash Duration :1 to 2000 seconds
Contact Outputs, controller :1 SPDT, 3 SPST relays
Switching Voltage / Current / Power :Max 250 VAC / Max 3A / Max 600VA
Function (Switchable) :Latching / pulse
Pickup Delay :0 to 2000 seconds
LCD :UV coat, backlit 14 segments display with symbols for status information
Backlight :On/Off selectable with four levels of brightness control
Emitted Interference :According to EN 61326
Immunity to Interference :According to EN 61326
Ambient Temperature Operating Range :0 to 50 °C
Maximum Relative Humidity :80% up to 31 °C decreasing linearly to 50% at 40 °C
Input :80 to 250 VAC/DC 50/60 Hz Approx. 10 VA
Main Fuse :Galvanic: 315 mA Anti-surge, 250 V Amperometric: 250 mA, Anti-surge
Pollution Degree :2
Transient Overvoltage Category :II
Transmitter Function :Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated
Load :Max 600 W
DO Input :Screw terminals
Connection Terminal (Galvanic) :5-pin, 8-pin, 9-pin and 13-pin terminal, detachable blocks
Connection Terminal (Amperometric) :Wall mount: 2x3-pin, 8-pin 9-pin and 13-pin terminal blocks Panel mount: 3-pin, 9-pin and 19-pin terminal blocks
Dimensions (W x H x D) :

Wall-mount (Galvanic): 144 x 144 x 111.5 mm

Wall-mount (Amperometric) : 144 x 144 x 110 mm

Panel-mount: 175 x 96 X 96 mm

Weight :

Wall-mount: 745g (Unit) / 1100g (Packed)

Panel-mount: 700g (Unit) / 850g (Packed)

Environmental Rating :

Wall-mount: IP66 (NEMA 4X)

Panel-mount: IP54 (front panel)