H2S - SO2 Analyzer, Model 450i

  H2S - SO2 Analyzer, Model 450i



Thermo Scientific's Model 450i Analyzer, catalytically converts Hydrogen Sulfide to Sulfur Dioxide and is then measured using pulsed fluorescence technology. The simultaneous measurement readings of H2S and SO2 provide reliable data capture on a constant basis. 


   Product Detail


Thermo Scientific's Model 450i  Hydrogen Sulfide/Sulfur Dioxide analyzer provides increased detection specificity and long term stability. 

Pulsed fluorescent technology is used to detect SO2 molecules that are formed by the catalytic conversion of H2S. SO2 molecules are excited by the Ultra Violet light and then decay to a lowered energy state producing a characteristic light that is measured by a photomultiplier tube. The pulsing of the UV source lamp increases the optical intensity through which greater U.V. energy throughput and lower detectable SO2 concentrations are realized. Reflective bandpass filters on the model 450i are more selective in wavelength isolation minimizing photochemical degradation.

 Key Features:
 • Pulsed excitation and intensity stabilization

 • Reflective bandpass filters

  Temperature stabilized reaction chamber

  Stable, long-life UV source
  Superior zero stability

Zero Noise 

 1.0 ppb
Manual H2S or SO2

 3.0 ppb Auto
(10 second averaging time) 

 0.5 ppb  1.5 ppb       
(60 second averaging time)
 0.25 ppb  0.75 ppb     
(300 second averaging time)

Lower Detectable Limit  

 2.0 ppb
Manual H2S or SO2
 6.0 ppb Auto
(10 second averaging time)
 1.0 ppb  3.0 ppb       
(60 second averaging time)
 0.5 ppb  1.5 ppb       
(300 second averaging time)  
 
Purchase Details
SKU : EID-10138850 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Preset Ranges :0-0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 ppm 0-0.2, 0.5, 1, 2, 5, 10, 20, and 25 mg/m³
Extended Ranges :0-0.5, 1, 2, 5, 10, 20, 50, and 100 ppm 0-2, 5, 10, 20, 50, 100, 200 and 250 mg/m³
Custom Ranges :0-0.05 to 100 ppm 0-0.2 to 250 mg/m³
Zero Drift (24 Hour) :Less than 1 ppb
Span Drift (24 Hour) :+/- 1% Full Scale
Response Time :80 Seconds (10 second average time) 110 Seconds (60 second average time) 320 seconds (300 second average time)
Precision :1% of reading or 1 ppb (whichever is greater)
Linearity :+/- 1% Full Scale
Sample Flow Rate :1.0 liter/min.
Converter Efficiency :> 80% H2S to SO2 (Note: various other sulfur compounds can be converted at varying %)
Operating Temperature :20°C - 30°C
Power Requirements :100 vac, 115 vac, 220-240 vac +/- 10% @ 300W
Size and Weight :16.75"(W) x 8.62"(H) x 23"(D), 48lbs. (21.8 kg)
Outputs :Selectable Voltage: RS232/RS485, TCP/IP, 10 Status Relays, and power fail indication (standard). 0-20 or 4-20 mA Isolated current output (optional)
Inputs :16 Digital inputs (standard), 8 0-10vdc Analog inputs (optional)