Ammonia Analyzer, Model 17i

  Ammonia Analyzer, Model 17i



Chemiluminescent Technology
NH3, NO, NO2, NOx, Nt
Designed for ambient air monitoring this i Series analyzer provides long term stability and high reliability.


   Product Detail


Using the light producing reaction of nitric oxide (NO) with ozone (O3) as the basic principle, Model 17i operates in three modes: NO, NOx, and Nt

          I. NO Mode: Sample is mixed with Ozone in the reaction chamber. This reaction produces a characteristic luminescence with intensity proportional to the concentration of NO.

II. NOx Mode: Sample is passed through a molybdenum converter to reduce any NO2 in the sample to NO. The sample is then transported to the reaction chamber where it is measured as NOx (NO + NO2 )

III. Nt Mode: Sample is passed through stainless steel converter where both the NO2 and the NH3 are converted to NO. The sample is then measured as Nt (NO + NO2 + NH3 )













 

 


 

The software subtracts NO from NOx and NOx from Nt and provides outputs of NO2 and NH3 respectively. The Model 17i can output NH3 along with NO, NO2, NOx, and Nt to the display or electronic outputs.

Key Features:
 • Measurement of NH3 as well as NO, NO2, NOx, and Nt
 • Replaceable NO2 and NH3 converter cartridges
 • Unparalleled sensitivity and selectivity
 • Automatic and manual modes
 • Selectable time constants
 
Purchase Details
SKU : EID-10141650 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Preset Ranges :0 - 0.05, 0.1, 0.2, 0.5, 1 ,2 ,5, 10, 20 ppm 0 - 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 30 mg/m³
Extended Ranges :0 - 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100 ppm 0 - 0.5, 1, 2, 5, 10, 20, 50, 100, 150 mg/m³
Custom Ranges :0 - 0.05 to 100 ppm 0 - 0.1 to 150 mg/m³
Zero Noise :0.50 ppb RMS (120 second averaging time)
Lower Detectable Limit :1.0 ppb (120 second averaging time)
Zero Drift (24 hour) :< 1 ppb
Span Drift (24 hour) :+/- 1% full scale
Response Time (0 - 90%) :120 seconds (10 second averaging time)
Precision :+/- 0.4ppb (500 ppb range)
Linearity :+/- 1% full scale
Sample Flow Rate :0.6 liters/min.
Operating Temperature :15°C - 35°C (59°F - 95°F)
Power Requirements :100 vac, 115 vac, 220-240 vac +/-10% @ 50/60hz, 300W (Analyzer) 600W (Converter)
Physical Dimensions :16.75" (W) x 8.62" (H) x 23" (D)
Weight: Analyzer :60lbs (28kg)
Weight: Converter :29lbs (14kg)
Outputs :Selectable voltage, RS232/RS485, TCP/IP, 10 status relays and power fail indicationn (standard). 0-20 or 4-20 mA isolated current output (optional)
Inputs :16 digital inputs (standard), 8 0-10vdc analog inputs (optional)