Welcome Guest from United States
Sign In Change Country
  0 Items
Search:
Product Browse Products >>... >> Air Quality Monitoring Solutions >> Source Gas Analyzers
Direct Methane, Non-Methane Analyzer, Model 55i

  Direct Methane, Non-Methane Analyzer, Model 55i



The Model 55i's measurement of methane and non-methane hydrocarbons (NMHC's) is based on the well established method of gas chromatography, utilizes flame ionization detection and features a proprietary column design to achieve ultimate separation of methane from all C2 compounds.


   Product Detail


    The Model 55i direct methane, non-methane hydrocarbon analyzer possesses unsurpassed sensitivity, repeatablity and specificity in ambient and source emissions. Configured for continuous, fully automated, stand-alone operation resulting in precise measurements.

The back-flush gas chromatography (GC) system is designed for automated measurement of methane and non-methane hydrocarbons. The GC allows for reliable measurement at sub-ppm concentrations, even in the presence of much higher methane concentrations. In addition, the proprietary column design is unaffected by oxygen, provides complete recovery of low volatility compounds and can achieve complete seperation of methane from all C2 compounds. 

A sample is transported through an eight port valve and detector oven. Based on the chemical and physical charateristics, methane will move at the highest velocity and emerges first.   The flame ionization detector will then measure the direct methane. When the eight port valve returns to the original positioning the back-flush of non-methane hydrocarbons will be activated for accurate reporting.

Key Features:

• Adjustable Ranges

• Real-time correction of THC readings 

• Automatic flame sensing and ignition

• Automatic calibration and span check

• 70 Second Analysis  

 
Purchase Details
SKU : EID-10142618 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Preset Ranges :0-5, 50, 500ppm, or 0-10, 100, 1000ppm, or 0-20, 200, 2000ppm, or 0-50, 500, 5000ppm
Zero Noise :0.025 ppm R
Minimum Detectable :0.0500 ppm CH4 - 0.050 NMHC as propane (300 second averaging time)
Zero Drift :Auto-zero each cycle
Span Drift (24hr) :< 2% span value (without auto-calibration)
Response Time (90%) :~ 70 Seconds
Precision :2.0% of reading or 50 ppb (whichever is greater)
Linearity :+/- 2% of span (at concentrations of 10% - 100% of span)
Sample flow rate :0.5 lpm, minimum
Makeup Air Flow Rate :275ccm - 350ccm Hydrocarbon free air
Fuel Flow Rate :25ccm to 50ccm Hydrogen or ~100-150ccm H2/He mixture
Carrier Gas Flow Rate :35ccm - 70ccm Nitrogen
Operating Temperature :+15°C to +35°C
Power Requirements :100vac, 115vac, 220-240 vac, +/-10%, 50/60Hz, 420W
Size and Weight :16.75" (W) x 8.62" (H) x 23" (D), 60lbs (17.7kg)
Outputs :Selectable voltage, RS232/RS485, TCP/IP, 10 Status relays and power fail indication. 0-20 or 4-20 mA Isolated current output (Optional)
Inputs :16 Digital Inputs (standard), 8 0-10vdc Analog inputs (optional)