FHT 750/751/752 BIOREM Neutron Detector


These series of neutron dose rate detectors feature excellent energy response and very low gamma interference. Among the three different versions available, a light-weight moderator has been designed for security applications.
   Product Detail


BIOREM  FHT 750 is a neutron dose rate detector for stationary and portable use, featuring excellent energy response according to ICRP 60 with special moderator.

It is currently available in three versions:

  • FHT 750 with built-in amplifier and external high voltage supply
  • FHT 751 with completely built-in microprocessor electronic with serial RS-485 interface for networked area monitoring 
  • FHT 752 for connection to Radiameter FH 40 G-L and FHT 6020

For enhanced neutron sensitivity we also build FHZ 752 H incorporating a He-3 tube as detector instead of a BF3 tube:

Neutron response: 2 cps at 1 µSv/h
Gamma response: 1 cps at 10 mSv/h

 
Purchase Details
SKU : FHT 750/751/752 BIOREM 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Energy Range :0.25 eV ... 10 MeV
Typical Neutron Response :1cps at 1.8 µSv/h
Gamma Rejection :10^-5 at 1 Sv/h due to BF3 detector
Background :typical 0.001 cps