Designed to continuously monitor gaseous iodine-131 in ambient air or effluent gas, the FHT 1700 boasts excellent selectivity against other iodine isotopes or interfering noble gases and enjoys a maintenance-free long-term operation. The gas ring vacuum pump is also maintenance-free and low noise. Operation is user-friendly with measuring electronics operable under Windows (TM).
Product Detail
The FHT 1700 continuously monitors gaseous 131I in ambient air or effluent gas:
Application in ambient air monitoring networks or stack emissions monitoring systems
Excellent selectivity against other iodine isotopes or interfering noble gases
Maintenance-free long term operation only limited by exchange of iodine adsorbent
Direct read-out of measured values in physical units of Bq/m3
Easy systems integration via RS232 serial interface
LAN coupling optionally available
Features
Measuring range for iodine concentration 1 Bq/h/m3 - 0.5 x 106 Bq/h/m3
Detector assembly comprising Marinelli beaker with iodine adsorbent and NaI(Tl) scintillation detector
Aerosol filter
Heated air inlet for optimised iodine retention
Measuring electronics FHT8000 operable under WindowsTM
Menu driven calibration procedure with complete track record of efficiency data
Low noise, maintenance free gas ring vacuum pump; air flow app. 5 m3/h.