The 2670 Farmer Dosemeter offers excellent stability and easy operation via large, clear menus. The instrument offers precise measurements of radiological beams with a choice of operational software.
Product Detail
The 2670 features traditional qualities with excellent performance achieved by state-of-the-art technology. In use, all menus, options and values are selected by two rotary knobs, giving a real "hands-on" feel to the instrument. A Chamber library and a Type library are included within the Calibration Mode, whose settings are protected by a keyswitch. The NE2670 measures dose, doserate, charge or current and includes a data link for 2-way communications with an external computer.
The 2670 is available in 2 variants:
The 2670A features a serial control option, for control and data collection by an external computer.
The 2670B features enhanced background calculation and supervisor options.
Please contact our sales staff for more information.
Features:
Intuitive control
Intrinsically stable Head Amplifier unique to us
Fast, manually-triggered autozero
Very short warmup and settling times
Repeatability: +0.05% of reading
Temp. Stability: <= 100 ppm per °C
EEPROM data store - no RAM battery required
Complies with IEC-731 & BS EN 61010-1
Frequently asked questions
PRODUCT#
Price
Features:
2670A
-
Serial link for computer control and data logging
SELECT
2670B
-
Advanced background correction and enhanced supervisor control
SELECT
Purchase Details
Contact Sales
United States +1 800 274 4212
Canada +1 800 274 4212
Caribbean +1 800 274 4212
Mexico +1 800 274 4212
Argentina +1 800 274 4212
Chile +1 800 274 4212
Other Latin Amer +1 800 274 4212
Austria +49 9131 998 226
Belgium +49 9131 998 226
Denmark +49 9131 998 226
Finland +49 9131 998 226
France +49 9131 998 226
Germany +49 9131 998 226
Ireland +49 9131 998 226
Italy +49 9131 998 226
Luxembourg +49 9131 998 226
Netherlands +49 9131 998 226
Norway +49 9131 998 226
Portugal +49 9131 998 226
Russia +49 9131 998 226
Spain +49 9131 998 226
Sweden +49 9131 998 226
Switzerland +49 9131 998 226
United Kingdom +44 118 971 5042
Other Europe +49 9131 998 226
China +49 9131 998 226
India +49 9131 998 226
Japan +49 9131 998 226
Middle East +49 9131 998 226
Other Asia +49 9131 998 226
South Africa +49 9131 998 226
Other Africa +49 9131 998 226
Australia +49 9131 998 226
Oceania +49 9131 998 226
Specifications:
Measured quantity: : Dose, Dose Rate, Charge or Current
Input current ranges: : 0 to 500 pA and 0 to 50 nA
Main display: : 5 digit floating point
Input resolution: : 0.001% of range full scale input
Input response time: : < 3s (to within 90% of final reading)
Non-linearity: : within +/- 0.03% of range full scale input
Repeatability: : +/- 0.05% of reading
Leakage current: : Head amplifier: 1 x 10^-15 A (typ)
With typical Farmer chamber attached:
- Range 1: 10 x 10^-15 A (typ)
- Range 2: 100 x 10^-15 A (typ)
Long term stability: : +/- 0.2% per year
Temperature stability: : <= 100 ppm per °C
Units: : Gy, Sv, R, rad, rem, C/kg, A, C
(prefixed by p, n, µ, m, c, k, M, G, T)
Calibration accuracy: : Charge: +/- 0.5% of uncorrected indicated value
Dose (with 0.6cm3 chamber): +/-0.5% against an NPL Secondary Standard
Autozero: : 60 seconds
Display: : 320 x 240 LCD graphics panel
Timer range: : 999999.9 s
Timer resolution: : 0.1 s
Timer accuracy: : 0.01%
Polarising voltage: : +/-30V to +/-1000V
Accuracy: +/-1% of reading, or +/-1V (whichever is the greater)
Temperature stability: : <=100 ppm per °C
Trip output: : changeover relay contacts for Start, Stop and Trip
Trip connector: : 9-way D socket
Serial link: : RS-232 via 9 way D socket (note: external computer control only available on 2670A)
Temperature: : +10 to +40°C operational. -15 to +40°C storage
Humidity: : up to 95% non-condensing
Power requirements: : 85 to 264 V AC, 47 to 440 Hz VA
Weight: : 6 kg nett