Dust Monitor, Model ADR-1200S


The Model ADR-1200S is a real-time particulate monitoring system designed for outdoor operation.  The unit is designed for continuous unatttended monitoring with continuous real-time data transmission to a central location and/or internal data logging.  Its weatherproof enclosure ensures safe an effective operation under a wide range of ambient environmental conditions.
   Product Detail


The ADR-1200S measures mass concentrations of particulates by means of light-scattering photometry.  The system has the capability to collect sampled particles on a membrane filter for gravimetric or chemical analysis.

The ADR-1200S can be used for particulate size-selective measurements using either the metal cyclone (included) or the optional inline impactor Model DR-PM10/2.5.  The latter is especially designed for PM10 and PM2.5 monitoring.  The cyclone is preferred for either respirable particle monitoring or to obtain information on particle size distribution.  In addition to the real-time/continuous measurements by light scattering, the system enables the user to collect the sampled particles on a membrane filter for gravimetric and/or chemical analysis.

Features:

  • Long-term, precise/driftless measurement of airborne particulate concentrations down to 1mg/m3
  • Real-time and weighed average measurements or particulate size selective measurements with appropriate aerodynamic perticle discriminators
  • On-board large-scale data logging (up to 13,000 data points in 99 discrete tags)
  • Data transmission with user-supplied modem or telemetry equipment
  • Digital (RS232), dual analog (4-20 mA and 0-5 V), and switched alarm outputs
  • NEMA-4 weatherproof enclosure can be pole-mounted, tripod-supported, or used on a benchtop
  • AC line-powered with 36-hour battery backup

Applications:

  • Hazardous remediation monitoring
  • Fenceline monitoring
  • Ambient monitoring
  • Roadside monitoring
  • Construction/demolition monitoring
  • Steel structure/bridge painting

 

 
Purchase Details
SKU : ADR-1200S 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Concentration measurement range (auto-ranging) :0.001 to 400 mg/m³
Precision/repeatability (2-sigma - at constant temperature) :+/-5 µg/m³ for 1 sec. averaging; +/- µg/m³ for 10-sec. averaging
Accuracy :+/-5% of reading +/- precision
Resolution :0.1% of reading or 0.001 mg/m³, whichever is larger
Particle size range of maximum response :0.1 to 10 µm
Concentration display updating interval :1 sec.
Alarm level adjustable range (user selectable) :selectable over entire measurement range
Alarm averaging time (user selectable) :real-time (1 to 60 sec.) or STEL (15 min.)
Data logging averaging periods (user selectable) :1 sec. to 4 hrs.
Total number of data points in memory :13,000
Number of data tags :99
Readout display :LCD 16 characters (4mm height) x 2 lines
Serial interface :RS232, 4800 baud
Analog signal output :0 to 5V and 4 to 20 mA, with selectable full scal ranges between 0.1 and 400mg/m³
Power :@115 VAC, 50 to 300 mA; @9VDC, depending on flow rate, flow load, and analog output current (if used)
Operating environment :14°F to 122°F (-10°C to 50°C), 10 to 95% RH, non-condensing
Storage environment :-4°F to 158°F (-20°C to 70°C)
Dimensions :6.9 in. (175 mm) D x 12 in. (305 mm) W x 16 in. (405 mm) H
Weight :22lb. (10kg) (without mounting brackets)
Enclosure rating :NEMA 4