Mini 6-96 Dual Scaler Ratemeter


The large display on the Mini 6-96 counter facilitates easy reading. A choice of inputs can be configured by the user in either preset or manual counting times.
   Product Detail


The 6-96 is a special purpose dual-scaler intended for use with alpha and beta counting drawer assemblies.

Features

  • Two six-digit crystal displays showing scaler count on 1cm high characters for ease of readability
  • Inputs can be configured for SC100 or 0744/0771/2 series drawer assemblies
  • Optional interface adapter for use with G-M tube or scintillation detectors
  • Pre-set counting times from 10 seconds to 1,000 seconds in seven steps plus manual timing using the start and top buttons
  • Mains operation with automatic switching for voltage supplies from 100 to 240V AC
  • Stabilized +15V supplies for powering drawer assembly pre-amplifiers
  • Tough, splashproof polycarbonate case

The 6-96 is intended for use with AEA alpha and beta counting drawer assemblies.   Interface cables are available for the 0744/0771/2 and SC100 series drawers.  An interface module is also available for connecting the scaler to a counting assembly where the detector is an end-window G-M tube or photomultiplier tube.  The only external controls are the start/stop buttons and time select switch, counts are shown on two 6-digit crystal displays.

 

Please note - The last day to purchase a Mini 6-96 is January 1, 2009.  Support will be available for this product until January 1, 2012.  The recommended replacement products are either the Delta 5 Benchtop/Portable Ratemeter or the RadEye B20 Lab1.

 
Related Products
Delta5 Benchtop/Portable Ratemeter
 Product#  
   
 Delta5 Benchtop  
SELECT   
 
 Product#  Product Name Image  
         
 RadEye B20 / RadEye B20-ER   RadEye B20 / RadEye B20-ER 
  SELECT  
 
Purchase Details
SKU : Mini6-96 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Input Options :SC100 input: (TTL pulse)
0744/0771/2 input: 12V, positive going to 0V on detector pulse
Drawer Power Supply :15V, 200 mA (SC100) or -15V 200 mA (0744/0771/2)
Detector Bias :-15V (optional 48V bias supply available to order)
Digital Scaler Display :2x6 digit LCD, 1 cm character height
Maximum count 899,000
Count overflow and count in progress indication
Timer :Crystal derived (163.84kHz) clock, accuracy +50 ppm
User selectable times of 10, 30, 60, 100, 300, 600, 1,000, and infinity
Paralysis Time :Internally set to 10 us
User Controls :Rotary on/off and combined time select. Push buttons for start/rest and stop
Power Supply :100 to 240V AC, 50-60 Hz, 15W. Automatic switching for input voltages
Housing & Dimensions :Polycarbonate case with a carrying handle and tilt feet
Size: 240 mm wide x 120 mm high x 100 mm deep
Weight :2 kg
Temperature Range :j-10oC to +45oC
Connectors :2 x "D" 9 pin connectors on rear of case for drawer interface cable
G-M Tube Interface, Type 696G1 :
Output Voltage :700 to 1,800V DC up to 50 uA
Input Sensitivity :1 pC (charge sensitive)
Power Supply Input :15V derived from dual scaler
Output Pulse :+5V, will drive one TTL load
Overload Indication :LCD on host 6-96 will blink at 1 Hz if the overload circuit trips
Housing & Dimensions :Painted aluminum case
Size: 240 mm wide x 12 mm high x 100 mm deep
Weight :1 kg
Temperature Range :-10oC to +45oC
Connectors :Probe: PET Series 100, 6-96 interface: 9 pin "D" type on 0.5 m cable
EMC/ESD :Both units comply with BS EN 50081-1 & BS EN50082-1 (severity level 2V/m). Electrical safety to IEC1010