ELEMENT GD

  ELEMENT GD



The Thermo Scientific ELEMENT GD, a combination of a glow discharge ion source with a high resolution mass spectrometer, is the ultimate tool for the direct analysis of conductive material. Almost all elements present in a solid sample, including carbon, oxygen and nitrogen can be detected and quantified routinely – many down to the parts per billion (ppb).

   Product Detail


  • The ELEMENT GD is tailored to serve high purity metal manufactures as well as their consumers from industries such as:
  • Microelectronic
    - Copper, Aluminum, Sputter Targets
  • Aerospace
    - Aluminum, Stainless Steel, Titanium Alloys
  • Medical/Pharmaceutical/Food
    - Stainless Steel, Titanium Alloys
  • Nuclear
    - Uranium, Nuclear Fuel

Have a look at our newest application report:
AN30164 AN30164: Analysis of Solar Cell Silicon using Glow Discharge Mass Spectrometry
Joachim Hinrichs, Meike Hamester, Lothar Rottmann,
Thermo Fisher Scientific, Bremen, Germany
=> Visit the Trace Elemental Home Page

ELEMENT GD Glow Discharge Ion Source

With the ELEMENT GD Thermo Electron combines components of superior attributes in an instrument incomparable in speed, sensitivity and accuracy.


Overview

  • Fast Flow Glow Discharge cell
    - Shortest analysis time due to high sputter rates
    - High power ion source for uppermost sensitivity
  • State of the art double focusing mass spectrometer
    - Signal to noise ratios enabling sub-ppb detection limits based on high ion transmission combined with low background.
    - Maximum level of selectivity and accuracy due to high mass resolution: a prerequisite for undisputable analytical results
  • 12 orders of magnitude automatic detection system
    - Determination of ultra-traces and matrix elements within a single analysis, due to a fully automatic detector with 12 orders linear dynamic range
    - Direct determination of the matrix elements for IBR (Ion Beam Ratio) quantification
  • State-of-the-art software suite for productivity and ease-of-use
    - Full computer control of all parameters
    - Fully automated tuning, analysis and data evaluation
    - Automatic LIMS connectivity
    - Remote control and diagnostic
    - Windows XP OS


  • Quantitative multi-element analysis across the periodic table
  • Less than 10 minutes sample turn-around
  • Matrix to ultra-trace detection capabilities in a single analysis
  • Depth profiling
  • Minimum matrix effects for straightforward quantification


The ELEMENT GD ion source and sample holder is designed for easy sample changeover, offering routine operation and high throughput.

Overview
The ingenious and yet simple design of the Glow Discharge Source reduces sample to sample switching to a fraction of the time it used to be. The sample itself is placed in a vacuum chamber, completely eliminating any risk of leaking between the sample and the GD cell (Figure 2). The sample holder can quickly and easily be removed from the ion source, simply unloaded and re-loaded, and repositioned for the next measurement (Figure 3). Opening and closing of the ion source manifold is fully automatic.

Please download the Winterplasma Conference 2009 Presentations

 

 
Verwandte Produkte
 PRODUCT# Product Name Image  
 
IQLAAEGAAM
FABWMAFC
 
ELEMENT 2
  Bitte wählen Sie  
 
IQLAAEGAAM
FABWMAFB
 
ELEMENT XR
  Bitte wählen Sie  
 
Purchase Details
SKU : IQLAAEGAAMFABWMAFA 
Request Quote
Contact Sales
 Sales
  Service
 
Specifications:
Sensitivity
(medium resolution) :
1x1010 cps
(Copper matrix)
Dark Noise :< 0.2 cps
Dynamic Range :> 1012 linear
with automatic gain calibration
Mass Resolution :3 fixed resolutions:
> 300; > 4.000; > 10.000
Resolution Switching Times :< 1 s
Mass Stability :25 ppm / 8 hour
Scan Speed (magnetic) :m/z 7 to 240 to 7 < 150 ms
Scan Speed (electric) :1 ms/jump, independent of mass range
Power :3-phase, 230/400 V ± 10 %, 50/60 Hz fused
32 A per phase
Environment :Temperature 18 - 24 °C (64 - 75 °F)
Humidity :50 - 60 %, non-condensing, non-corrosive
Cooling Water :∼200 L/h
Temperature 10 - 20 °C
4 - 6 bar (43 - 65 psi)
Argon :1 L/min Argon 5.0 or better.
5 L/min Argon 4.6
Regulated pressure :8 - 10 bar (116 - 145 psi)
Pump Exhaust :1 x 25 mm
Flow rate :1 L/min
Electronic Exhaust :1 x 15 cm , 400 m3/h