FHT 192

  FHT 192



The FHT 192 Ionization chamber can be connected to either the FHT 6020 area monitor or the FH 40 G survey meter, for measurements at high energy accelerators or X-ray sources

   Product Detail


The new FHT 192 is based on the famous PTB-approved medium-pressure chamber FHT 191 N, which is used as a reference standard detector by many organizations. The chamber geometry design offers both excellent angular dependence (- 45° to +90°) and energy dependence. The chamber is equipped with a styrofoam protection in order to minimize temperature change induced currents.

Features:

? 30 keV to 7 MeV energy range

? 100 nSv/h to 1 Sv/h dose rate range

? ¾ sphere angular response:

? integrated preamplifier fits to
  ? FHT 6020 area monitors
  ? FH 40 G survey meter
  ? SVG2 military Radiac

? high dose option with remote preamplifier

 
Produits associés
 PRODUCT# Product Name Image  
  FH 40 G   FH 40 G Multi-Purpose Survey Meter
  Choisir  
  FHT 6020   FHT 6020 Area Monitor Display and Alarm Unit
  Choisir  
  SVG2   SVG2 RadiacMeter
  Choisir  
 
Purchase Details
SKU : FHT 192 
Request Quote
Contact Sales
 Sales
  Service
 
Specifications:
Detector type :Pressurized ionization chamber
Measured variable :Ambient equivalent dose rate H*(10)
Measuring range :100 nSvh-1 - 1 Svh-1
Energy range :30 keV - 7 MeV (± 30%)
Gamma- responsiveness :250 fA/µSvh-1
Angular acceptance :-45° < F < +90° (270° solid angle)
Overload :Up to 50 Sv/h overload indication
Operating gas (permanently filled) :Noble gas - nitrogen - mixture
Filling pressure :7000 hPa (absolute value)
Volume :4.25 liters