  Model 43i- HL High Level SO2 Analyzer



A change is in the air!

The industry’s new best of breed. Our customers told us exactly what they’re looking for in a gas monitoring solution: reliability, simplicity and ease of use. The new iSeries platform delivers on all counts- and than goes a step farther. A major product in Thermo’s new iSeries product line is the Model 43i-HL SO2 analyzer.

   Product Detail


Using pulsed florescence technology, the Model 43i-HL SO2 measures th amount of sulfur dioxide in the air from sub ppm levels up to 1,000 ppm in standard range mode and from 100ppm up to 10,000ppm in extended range.

Dual range and auto range are standard features in this instrument as well as temperature and pressure correction. User settable alarm levels for concentration and for a  wide variety of internal diagnostics are available from an easy to follow menu structure.

This state of the art gas analyzer offers features such as an ethernet port as well as flash memory for increased data storage and field upgradeability.

Ethernet connectivity provides efficient remote access, allowing the user to down load measurement information directly from the instrument without having to be on-site.

You can easily program soft-keys to allow you to jump directly to frequently accessed functions, menus or screens. The larger interface screen display measurement information and status, while viewing menu and operational screens.

Key Features

  • Ethernet connectivity for efficient remote access
  • Enhanced user interface with one button programming and large display screen
  • Flash memory for increased data storage and user downloadable software
  • Enhanced electronics design optimizes product commonality
  • Improved layout for easier accessibility to components

Comprehensive Service Solutions

To maintain optimal product performance, you need immediate access to experts worldwide, as well as priority status when your air quality equipment needs repair or replacement.  Thermo Scientific offers comprehensive, flexible support solutions for all phases of the product lifecycle.  Through predictable, fixed-cost pricing, Thermo services help protect the return on investment and total cost of ownership of your Thermo Scientific air quality products.

 
Related Products
 Product#  Product Name Image  
 43i-TLE   Enhanced Trace Level SO2 Analyzer, Model 43i-TLE 
  Select  
 43i   SO2 Gas Analyzer, Model 43i 
  Select  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Firmware Downloads for Air Quality Instruments
  • Software Downloads for Air Quality Instruments
  • Specifications:
    Standard Ranges :0-10, 20, 50, 100, 200, 500 and 1000ppm
    0-20, 50, 100, 200, 500, 1000 and 2000 mg/m3
    Extended Ranges :0-100, 200, 500, 1000, 2000, 5000 and 10000ppm
    0-200, 500, 1000, 5000, 10000 and 20000 mg/m3
    Custom Ranges :0-10 to 1000 ppm (standard), 0-100 to 10000 ppm (extended)
    0-20 to 2000 mg/m3 (standard), 0-200 to 20000 mg/m3 (extended)
    Zero Noise :0.25 ppm (300 second averaging time)
    Lower Detectable Limit :0.5 ppm (300 second averaging time)
    Zero Drift (24 Hour) :less than 1ppm per day
    Span Drift (24 Hour) :+/- 1% full scale
    Response Time :80 seconds (10 second averaging time)
    Linearity :+/- 1% full scale
    Sample Flow Rate :0.6 to 1.0 liters/min
    Interferences (EPA levels) :< lower detectable limit except for the following:
    NO: < 3 ppb, tested at 500 ppb, M-Xylene <1 ppb, tested at 200ppb, H2O <2 % of reading, tested at 2% H2O absolute
    Operational Temperature :5° C- 40° C (20° C-30° C with EPA designation )
    Power Requirements :100 vac. 115 vac, 220-240 vac +/- 10% @ 300W
    Size and Weight :16.75" W x 8.62 H x 23" D, 55lbs / 425mm W x 219 H x 584mm D, 25kg
    Outputs :Selectrable voltage, RS232/ RS485, TCP/IP, 10 status relays, and power fail indication (standard)
    0-20 or 4-20 mA isolated current output (optional)
    Inputs :16 digital inputs (standard)
    8 0-10vdc analog inputs (optional)