Continuous Particulate TEOM Monitor,  Series 1405

  Continuous Particulate TEOM Monitor, Series 1405



The TEOM Series 1405 is a family of monitors that provide the opportunity to measure PM-2.5, PM-10, and PM-Coarse simultaneously or separately.

New! The Thermo Scientific Continuous Particulate TEOM Monitor, Model 1405-DF has been awarded US EPA PM-Coarse FEM designation: EQPM-0609-182. 
Details...

   Product Detail


Built on proven Tapered Element Oscillating Microbalance (TEOM) and Filter Dynamics Measurement System (FDMS) technologies, this line offers expanded capabilities, dramatically improved ease of use and a full spectrum of monitoring options. TEOMs provide NIST traceable direct mass measurements which circumvents uncertainties inherent in surrogate techniques such as beta attenuation, light scattering and pressure drop. The FDMS option provides the ability to account for the volatiles and semi-volatiles.

The 1405 series comes with new communication software for Ethernet connections, intuitive touch screens with calibration and maintenance tools, and a quick diagnostic system. The series also has the ability to be configured with or without the Filter Dynamics Measurement System (FDMS) in a single cabinet depending on site specific needs and monitoring requirements.

For varied site needs there are four models to choose from:

Model 1405
Single Flow TEOM
 Measures PM-2.5 or PM-10
Model 1405-D
Dual Flow TEOM
 Measures PM-2.5, PM-Coarse and PM-10
Model 1405-F
Single Flow TEOM with FDMS
 Measures PM-2.5, PM-10
and accounts for volatiles and semi-volatiles
Model 1405-DF
Dual Flow TEOM with FDMS
 Measures PM-2.5, PM-Coarse, PM-10
and accounts for volatiles and semi-volatiles

For more information on individual models please click on the appropriate datasheet fo the left.

 Continuous Particulate TEOM Monitor, Series 1405
 Product# Price  
     
 
Related Products
 Product# Product Name Image  
 
Streamline
Pro
 
Streamline Pro
  Select  

Alternatives
 
 Product# Product Name Image  
  1400ab   Continuous Particulate TEOM Monitor, Series 1400ab
  Select  
 
EID-101228
10
 
Filter Dynamics Measurement System, Model 8500
  Select  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Measurement Range :0 to 1,000,000 µg/m³
Resolution :0.1 µg/m³
Precision :±2.0 µg/m³ (1-hour ave), ±1.0 µg/m³ (24-hour ave)
Accuracy for Mass Measurement :±0.75%
Data Storage :Up to 500,000 records of user-specified variables
User Interface :Touch screen display, ePort software to view and change system operation from PC
Data Ouput and Input :Ethernet with embedded FTP server, USB, RS232, RS485, 8 user-defined analog outputs, 2 user-defined contact closure alarms circuits, 4 averaged analog inputs
Operating Range :Sampled air: -40 to 60 °C. Instrument must be weather protected within the range of 8 to 25 °C
Power Requirements :1405-DF Unit: 100-240 VAC, 440 VA, 47-63 Hz, Sampling pump, 120 VAC/60 Hz: 4.25A; 240 VAC/50 Hz: 2.25A
Weight :38 - 75 lbs (18 - 34 kg) depending