Harshaw TLD Model 3500, Manual Reader

  Harshaw TLD Model 3500, Manual Reader



The Harshaw 3500 TLD Reader provides cost-effective measurements of the radiation dose absorbed by individual TLD elements: ribbon (chips), rods, micro-cubes or powders.
   Product Detail


lIntroduction

The Harshaw 3500 operates on WinREMS Software, which runs under Windows® on a separate computer, providing user interface while the reader control and the applications software. The instrument includes a sample drawer for a single element TLD dosimeter, a linear, programmable heating system and a cooled photomultiplier tube with associated electronics to measure the TL light output. The manually operated Harshaw 3500 finds applications in medical physics and dosimetry laboratories.

Harshaw 3500 Key Features

  • Planchet heating incorporates welded thermocouple for best temperature reproducibility
  • Heating profile includes pre-heat and post-read anneal cycles, with editing
  • Heating temperature capability up to 600°C
  • 7 decade glowcurve acquisition range
  • Thermoelectric PMT cooler for maximum gain stability
  • Measurement quality assurance
  • Automatic background subtraction capability
  • Easy to operate, service and maintain
  • Compact and attractive (320 mm wide x 290 mm deep x 290 mm high)
  • Optional calibration software
  • Optional neutral density filter to extend the high measurement range

Advantages of a Separate Computer

  • Minimum initial investment
  • Extremely flexible parametric adjustments, implemented in software
  • The computer can be used for other purposes when not required for TLD
  • Use of commercial software for data manipulation, report generation and storage

Applications

  • Radiotherapy planning verification
  • Total Body Irradiation dose verification
  • Skin irradiation dose verification
  • Stereotactic beam output factor measurement
  • Critical organ dose verification
  • Diagnostic dose studies
  • CT dose measurement for quality assurance
  • Environmental dosimetry
  • Testing for irradiated food
  • Radioactive dating

Available Dosimeters

  • Recessed planchets for chip sizes
    3.2mm x 3.2mm, 6.4mm x 6.4mm, 15.7mm round
  • Chip and Rod planchet
    3.2mm x 3.2mm chip and 6mm rod, 1mm dia. or 1mm sq.
Accessories
  • Personal Computer
  • Printer
  • N2 gas flowmeter
  • 2210 Tabletop Irradiator
  • Low temp annealing oven
  • High temp annealing oven
  • Programmable annealing oven
  • Uninterruptable power supply
  • Vacuum tweezers
  • Dosimeter storage tray
  • Stainless steel annealing tray assembly
  • Stainless steel annealing tray

STANDARD FUNCTIONS PERFORMED BY WinREMS SOFTWARE

Graphical user interface

  • Pull-down menus
  • Selection via mouse


TL data acquisition parameter selections

  • Application of Reader Calibration Factor, (RCF), and Element Correction Coefficient, (ECC)
  • Automatic background subtraction
  • Application of quality factor
  • Raw data and glowcurve printing during acquisition
  • ASCII export file generation
  • Periodic PMT testing of noise and response to test light
  • Alarm and stop if TL data exceeds preset limit
  • Re-read TLD if TL data exceeds preset limit

TLD heating profile setup

  • Selection of pre-defined Time/Temperature Profiles (TTPs)
  • Editing of: TTP titles, Region of Interest (ROI) limits, Calibration region, Preheat temperature and time, Maximum read temperature, Temperature ramp rate, Anneal temperature and time, Acquisition time

TL data acquisition

  • Acquisition of glowcurve
  • Acquisition of temperature profile

TL data presentation and storage

  • Reading parameters
  • Date and time of reading
  • Dosimeter identification
  • Graphically presented glowcurve and TTP
  • Integrals of up to four regions of interest
  • Applicable RCF and ECC
  • Background value for subtraction
  • Values of test light and PMT dark current readings

Software options

  • Dosimeter calibration software
  • Glowcurve deconvolution

Computer requirements (minimum)

Pentium 166 MHz with 32 MB RAM; 40 MB free disk space; CDROM drive
SVGA color monitor (800 x 600), keyboard and mouse.
Windows™ 98, 2000, XP or NT 4.x
One RS232 serial port for connection to Model 3500 and one parallel port for printer.
 
Related Products

Accessories and Related Products
 
 Product#  Product Name Image  
 Harshaw TLD 90-Sr Irradiator option   Harshaw TLD 90-Sr Internal Irradiator (66 & 8800) 
  Select  
 Harshaw TLD 2210   Harshaw TLD Model 2210, Chip Irradiator 
  Select  
 Harshaw TLD 4488   Harshaw TLD Model 4488, Dry Air Generator 
  Select  
 Harshaw TLD 4500   Harshaw TLD Model 4500, Manual reader 
  Select  
 Harshaw TLD 5500   Harshaw TLD Model 5500, Automatic reader 
  Select  
 Harshaw TLD Model 6600 Plus   Harshaw TLD Model 6600 Plus 
  Select  
 Harshaw TLD 6610   Harshaw TLD Model 6610, Irradiator 
  Select  
 Harshaw TLD 8800   Harshaw TLD Model 8800, High Capacity Auto System 
  Select  
 
Purchase Details
SKU : 3500 TLD 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Radiations Measured :Photon, energies >5 keV; Neutron, thermal to 100 MeV; Electron/beta, energies >70 keV
Range :10 µGy to 1 Gy (1 mrad to 100 rad) linear; 1 Gy to 20 Gy (100 rad to 2000 rad) supralinear
Tissue Equivalence :nearly tissue equivalent
Fading :<20% in 3 months without thermal treatment; <5% in 3 months using preheat or glow curve deconvolution
Batch Uniformity :±15% std. dev. from the batch mean, tighter on special request
Residual TL Signal :<0.2% of reading, over the range, without annealing
Repeatability :for 1 mGy (100 mrad) 137Cs doses, <2% STD DEV of 10 sequential measurements
Reuse :more than 500 per dosimeter with <10% sensitivity change
Threshold :<10 µGy (1 mrad) based on 2.26 std. dev. of 10 repeat readings of an unexposed dosimeter