Welcome Guest from United States
Sign In Change Country
  0 Items
Search:
Product Browse Products >>... >> Online Electrochemistry >> Instruments/Analyzers
alpha-COND 1000 1/4 DIN Conductivity Controller

  alpha-COND 1000 1/4 DIN Conductivity Controller



The alpha-CON1000 conductivity controller offers ten different ranges of conductivity.  Conductivity ranges and its appropriate cell constant can be selected simply from the keypad through the setup menu.  No need to select jumpers!

 


   Product Detail


  • Ten selectable conductivity measurement ranges in one controller via its IP54 front panel. High-level accuracy of ±1% of full scale can be obtained with appropriate cells and correct temperature coefficient
  • User-Customization through Advanced Setup Menu offers flexibility in matching the controller's functions to suit individual's specific requirement
  • Choice of Cell Constant (0.01, 0.1, 1.0, 10.0) for accurate control in any solution
  • Hold Relay for use with float switches/flow switches and other controllers as a failsafe function
  • Two Level Password Protection prevents un authorized tampering with settings
  • 0 to 2000 Second Time Delay Adjustment on control and alarm delays
  • Galvanically Isolated Scalebale 0-20/4-20 mA Outputs
  • Wash Contact Relay controls electrodes cleaning systems at desired duration and frequency
  • Adjustable Hysteresis (Dead Band) prevents rapid contact switching near set point
  • Non-Volatile Memory retains all stored parameters and calibration data even if power fails
  • Line Resistance Compensation against intrinsic cable resistance for longer cable connection
  • Large Dual Display shows measurement with temperature simultaneously - features clear multiple icons, set points, and status messages
  • Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or 3-wire temperature input selection
  • Easy Installation and Wiring with detachable plug-in connectors

Applications

General: Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator.

Industrial: Use in applications involving agriculture, chemical processing, boiler and water heaters, wafer-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries, and bleach manufacturing.

Water Purification/Treatment: Use to treat batches of incoming process water, ultrapure water, boiler and feed water control.

Regulatory: Hook up to recorder for regulatory compliance data collection

 alpha-COND 1000 1/4 DIN Conductivity Controller
 Product# Price  Item  
       
 TSCONCTP1001      -    alpha COND 1000 Controller / Transmitter with 0/4-20mA output and 110VAC setting (unit only)   SELECT  
 TSCONCTP1002      -    alpha COND 1000 Controller / Transmitter with 0/4-20mA output and 220VAC setting (unit only)   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Conductivity Range :0.000 to 1.999 µS/cm 0.01 µS/cm 0.1 µS/cm 1 µS/cm 5 µS/cm 0.01 mS/cm 0.1 mS/cm 0.001 µS/cm 0.01 µS/cm 0.1 µS/cm 1 µS/cm 5 µS/cm 0.01 mS/cm 0.1 mS/cm
Accuracy :±1 % full scale
Temperature Range :-9.9 to 125 °C
Resolution & Relative Accuracy :0.1 °C & ±0.5 °C
Temperature Sensor :Pt 100/ Pt 1000 (jumper selectable)
Temperature Compensation :Auto / Manual (reference at 25 °C)
Function :Limit / Proportional (pulse or frequency) controller
Switching Hysteresis :0 to 10% full scale
Pickup / Dropout Delay :0 to 2000 seconds
Contact Outputs, Controller :2 potential free change-over contacts
Switching Voltage / Current / Power :Max 250 VAC / Max 3A / Max 600VA
Function (Switchable) :Launching or Pulse
Pickup Delay :0 to 2000 seconds
Switching Voltage / Current / Power :Max 250 VAC / Max 3A / Max 600VA
Power requirement :110/ 220 VAC (jumper selection)
Frequency :48 to 62Hz
Signal Output :0/4 to 20mA, galvanically isolated
Load :Max. 600 ?
Connection Terminal :Terminal blocks 5-pole/ 17-pole, removable
Main Fuse/ Fine Wire Fuse :Slow-blow 250 V/ 100 mA
Emitted Interference :According to EN 50081-1
Immunity to Interference :According to EN EN 50082-1
Maximum Relative Humidity :10 to 95%, non-condensing
Ambient Temp. Operating Range :0 to 50 °C
Dimensions :175 x 96 x 96 mm
Weights :Max. 0.7kg
Material :ABS with polycarbonate