Orion 2100R Industrial ROSS Ultra pH Electrode

  Orion 2100R Industrial ROSS Ultra pH Electrode



The Thermo Scientific Orion Process Ross Ultra offers the ultimate performance of a ROSS Ultra pH electrode technology combined in a rugged industrial housing offers the fastest most accurate response in a wide variety of samples.


   Product Detail


The 2100R series ROSS Ultra process electrodes are industrial combination pH electrodes designed for accurate pH measurements in the harshest environments and samples.  The 2100R series ROSS Ultra electrodes have a double junction, are non-refillable and include automatic temperature compensation and solution ground.

The four-pronged epoxy bulb guard protects the pH-sensing bulb from particulates or solids in a flowing sample.  The electrode has 3/4" NPT threads on both ends for mounting it within a pipe system or at the end of a conduit extension piece for submersion applications.  The 2100R series ROSS Ultra electrodes are compatible with the Thermo Scentific Orion and Alpha online pH monitors.

 Orion 2100R Industrial ROSS Ultra pH Electrode
 Product# Price  Description  
       
 2100R5      -    Process Ross Ultra flat bulb with PT 100 Thermistor and Industrial Housing. Electrode, with 5 meter cable   SELECT  
 2100R10      -    Process Ross Ultra flat bulb with PT 100 Thermistor and Industrial Housing. Electrode, with 10 meter cable   SELECT  
 21000R5      -    Process Ross Ultra flat bulb with PT 1000 Thermistor and Industrial Housing. Electrode, with 5 meter cable   SELECT  
 21000R10      -    Process Ross Ultra flat bulb with PT 1000 Thermistor and Industrial Housing. Electrode, with 10 meter cable   SELECT  
 2130R5      -    Process Ross Ultra flat bulb with 30K Thermistor and Industrial Housing Electrode, with 5 meter cable   SELECT  
 2130R10      -    Process Ross Ultra flat bulb with 30K Thermistor and Industrial Housing Electrode, with 10 meter cable   SELECT  
Purchase Details
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Specifications:
Applications :Drinking Water, Waste/Municipal, Industrial Water, Environmental
Length x diameter (mm) :160 mm x 33 mm
Reference System :ROSS Ultra combination with ultra high molecular weight polyethlyene (UHMWP) junction
Pressure Range :0 to 120 psi
Temperature Range :-5 to 100 ºC (32 to 212ºF)
pH Range :0 - 14
Drift :≤0.5 pH units for 30 days
Reproducibility :±0.01 pH
Slope :92 to 120% (54.4 to 60.3 mV/dec)
Reference Electrolyte :Non-refillable, ROSS polymer gel
Wetted Materials :Ryton body (glass-filled) with silicone front seal
Cable :Offered in 5 meter or 10 meter lengths, zero halogen low smoke integral coaxial cable terminated with stripped and tinned wires