  Model 42C High Level Analyzer - See Model 42i HL


Using Chemiluminescence, the 42C series is capable to measuring oxides of nitrogen from sub parts per billion (ppb) to 5,000 parts per million (ppm).  Extended troubleshooting diagnostics provide instantanuous indication of instrument operating status.  Reliable. Industry standard. U.S. EPA designated method RFNA-1289-074

This product is no longer offered.
   Product Detail


The Model 42C High Level (42-HL) combines the superior optical, mechanical, and chemical characteristics of its predecessor, the Model 42H, with an enhanced electronics package and user interface. The outcome is a powerful, easy-to-use, chemiluminescent based analyzer capable of measuring oxides of nitrogen from parts per billion (ppb) to 5000parts per million (ppm).

Thermo's design engineers have introduced a user interface which easily guides one through operation of the Model 42C-HL. This is accomplished via 4 line by 20 character Vacuum Fluorescent Display and simple layout of easy-to-use function keys. The end result is a combination of self explanatory display messages and intuitive function entries.

User programmable software capabilities allow individual measurement range settings to be stored in memory for subsequent recall and NO, NO2, NOX hourly average storage for up to one month. Extended troubleshooting diagnostics now provide instantaneous indication of instrument operating parameter status including: Pressure, Flow, DC Supply, Voltages, Internal emperate, Reaction Chamber Temperature, PMT Operating Voltage and Converter Temperature.


 KEY FEATURES

  • Wide Dynamic Range
  • Electronic Diagnostic Transducers
  • Multi-line Alpha Numeric Display
  • Dedicated Communications Processor
  • Remote Performance Diagnostics
  • Oil-Less Vacuum Pump
Related Products

 Product#  Product Name Image  
         
 42iHL   High Level NO-NO2-NOX Analyzer, Model 42i HL 
  SELECT  
 
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Software Downloads for Air Quality Instruments
  • Specifications:
    Range :0-10, 20, 50, 100, 200, 500, 1000, 2000, 5000 ppm
    Noise :25 ppb
    Detection Limit :50 ppb
    Zero Drift (24 hour) :50 ppb
    Span Drift (24 hour) :+/- 1% full scale
    Rise Fall Times (0-90%) :2.5 seconds NO Mode; 5.0 seconds NOX Mode
    Linearity :+1% full scale
    Sample Flow Rate :25 cc/min. with Bypass 2 SCFH
    Vacuum :28.5" Hg
    Power Requirements :90-110, 105-125, 210-250 VAC 500 Watts
    Physical Dimensions :17"(W) x 8 ¾" (H) x 23" (D)
    Weight :70 lbs., including pump
    Outputs :selectable voltage 0-10mv, 100mv, 1v, 5v, 10v, 4-20mA isolated (optional)