Model 43C SO2 Analyzer - See Model 43i

  Model 43C SO2 Analyzer - See Model 43i



Reflective UV filtering offers superior sensitivity with multiple-range settings from 0-5,000 ppm.  The 43C is a benchmark for sensitivity, stability, and selectivity.  With its U.S. EPA Designation of EQSA-0486-060, the Model 43C is an industry standard for SO2 monitoring.

This product is no longer offered.
   Product Detail


The Model 43C integrates the proven pulsed flurorescence design of Thermo Scientific's Model 43 series with an enhanced electronics package and user interface.  The outcome is a sensitive, ultra stable SO2 analyzer offering network operators and research scientists unlimited troubleshooting diagnostics and data communications capability.

User software facilities include field programmable measurement ranges and SO2 concentration value storage by date and time.

Extended troubleshooting diagnostics now provide instantaneous indication of instrument operation parameter status including Pressure Flow, DC Supply Voltage, Internal Temperature, Reaction Chamber Temperature, PMT Operating Voltage, Lamp Intensity, Lamp Voltage and Optical Span Test.

KEY FEATURES

  • Patented Pulsed Fluorescence             
  • Electronic Diagnostic Transducers             
  • Multi-Line Alpha Numeric Display             
  • Dedicated Communications Processor             
  • Remote Performance Diagnostics             
  • U.S. EPA Designated Method (EQSA-0486-060)
Related Products

 Product#  Product Name Image  
         
 43i   SO2 Gas Analyzer, Model 43i 
  SELECT  
 
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Software Downloads for Air Quality Instruments
  • Specifications:
    Ranges :0-50, 100, 200, 500, 1000 ppb and 0-1, 2, 5, 10, 20, 50, 100 ppm
    Zero Noise :1.0 ppb RMS (10 second averaging time); .05 ppb RMS (60 second averaging time); .25 ppb RMS (300 second averaging time)
    Lower Detectable Limit :21.0 ppb RMS (10 second averaging time); .05 ppb RMS (60 second averaging time); .5 ppb RMS (300 second averaging time)
    Zero Drift (24 hour) :less than 1 ppb
    Span Drift (24 hour) :±1%/week
    Response Time :80 sec (10 second averaging time); 100 sec (60 second averaging time); 320 sec (300 second averaging time)
    Precision :1% of reading or 1 ppb
    Linearity :+ 1% full scale
    Sample Flow Rate :0.5 liters/min (standard); 1 liter/min (optional)
    Operating Temperature :20° - 30°C (may be safely operated over the range of 5° - 45°C)
    Power Requirements :105-125 VAC, 60 Hz 220-240 VAC, 50 Hz; 100 Watts
    Physical Dimensions :16.75" (W) x 8.62" (H) x 23" (D)
    Weight : 44 lbs.
    Outputs :Selectable voltages and RS-232 (standard); 4-20mA isolated current (optional)