Model 48C Trace Level CO Analyzer

  Model 48C Trace Level CO Analyzer



For high sensitivity and selectivity in air monitoring, the 48C Series utilizes Gas Filter Correlation to measure high concentrations of carbon monoxide.  Combining proven detection technology and experience in the field, this series can detect CO levels from 0 to 20,000 ppm.

This product is no longer offered.

   Product Detail


For High Sensitivity Air Monitoring

The   Model 48CTL  Gas Filter Correlation (GFC) CO Analyzer measures low CO concentrations. It combines proven detection technology, easy to use menu-driven software, and advanced diagnostics to offer unsurpassed flexibility and reliability.

Thermo Environmental Instruments design engineers  introduced a user interface which easily guides one through operation of the   Model 48CTL. This is accomplished via a 4 line by 20 character Vacuum Fluorescent Display and simple layout of easy-to-use function keys.  The end result is a combination of self explanatory display messages and intuitive function entries.

The   Model 48CTL  is based on the principle that carbon monoxide (CO) absorbs infrared radiation at a wavelength of 4.6 micron.  Because infrared absorption is a nonlinear measurement technique, it is necessary for the instrument electronics to transform the basic analyzer signal into a linear output. The  Model 48CTL  uses an exact calibration curve to accurately linearize the instrument output over any range up to a concentration of 1,000 ppm.

The sample is drawn into the analyzer through the SAMPLE bulkhead. The sample flows through the optical bench. Radiation from an infrared source is chopped and then passed through a gas filter alternating between CO and N2 . The radiation the passes through a narrow bandpass interference and enters the optical bench where absorption by the sample gas occurs. The infrared radiation then exits the optical bench and falls on an infrared detector.

KEY FEATURES:

  • USEPA Designated Method (RFCA-0981-054)    
  • Gas Filter Correlation Selectivity    
  • Highly Specific to CO    
  • Electronic Diagnostic Transducers    
  • Multi-Line Alpha Numeric Display    
  • Dedicated Communications Processor    
  • Remote Performance Diagnostics    
  • Self Aligning Optics
  •  
    Related Products
     Product# Product Name Image  
      48C   Model 48C CO Analyzer - See Model 48i
      Select  
      48CHL   Model 48C High Level CO Analyzer - See Model 48i HL
      Select  
     
    Product Contact
     Sales Contact
      Service Contact
     
    Specifications:
    Range :0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000 ppm
    Zero Noise :0.01 ppm RMS for 300 second signal averaging time
    Minimum Detectable Limit :0.02 ppm
    Zero Drift (24 hours) :+/- 0.1 ppm
    Span Drift (24 hours) :+/- 2% Full Scale
    Rise, Fall Times (0-95% at 1 Lpm flow) :1 min. (30 sec. signal averaging time); 5 min. (300 sec. signal averaging time)
    Precision :+/- 1% Full Scale or +/- 0.02 ppm
    Linearity :+/- 1% of Full Scale or +/- 0.02 ppm
    Flow Rate :0.5 - 2 Lpm (1 Lpm is standard)
    Interference Rejection Ratios :H2O: 200,000:1 ; CO2: 80,000:1 (at 1,000 ppm CO2)
    Operating Temperature :Performance specifications maintained over the range 20-30°C (may be operated safely over the range 5-45°C)
    Power Requirement :105-125 Vac, 60 Hz; 220-240 Vac, 50 Hz, 100 Watts
    Physical Dimensions :17" (W) x 8-3/4" (H) x 23" (D)
    Weight :45 lbs
    Dual Outputs (Standard) :Individually selectable to 0-10mv, 0-100mv, 0-1V, 0-5V, 0-10V, digital display, 1 hour integrated value. Other outputs available upon request (4-20mA)