To meet the demands of the Gas Turbine CEMS market and other applications where medium range NO-NO2-NOX measurements are required, Thermo Fisher Scientific introduces the Model 42C Low Source Analyzer.  This instrument is specifically designed for optimal operation in the mid level ranges from 0.0-0.5 ppm to 0.0-200 ppm.



This product is no longer offered.
   Product Detail


The Model 42C-LS combines superior optical, mechanical and chemical characteristics of its predecessors with an enhanced electronics package and user interface.

Extended troubleshooting diagnostics now provide instantaneous indication of instrument operating parameter status including: Sample Pressure, DC Supply Voltages, Internal Temperature, Reaction Chamber Temperature, PMT Operating Voltage, and Converter Temperature.

The Use of chemiluminescence to detect NO/NOx has shown itself to be an excellent gas monitoring method. Using this underlying technology, TEI has developed a family of NO/NOx analyzers that can be tailored for many environments. However, the advent of gas-fired turbines places a new demand on NOx analyzers resulting in the need for an instrument optimized for 2  200 ppm and designed to operate within the constraints of the gas-fired turbine exhaust stream.


KEY FEATURES

  • Dual Range 0-2 and 0-200 ppm
  • Designed to meet EPA Low NOX Regulations
  • Optimized to minimize interferences
  • Long life stainless steel converter
  • Internal pump and proprietary ammonia scrubber for  SCR & SNCR applications
Related Products

 Product#  Product Name Image  
         
 42iLS   Low Source NO-NO2-NOX Analyzer, Model 42i LS 
  SELECT  
 
Product Contact
 Sales Contact
  Service Contact
 
Services
  • Software Downloads for Air Quality Instruments
  • Specifications:
    Ranges :0-2 ppm; 0-200 ppm (extended range)
    Custom Ranges :0-60 ppm; (0-200 ppm in extended range)
    Zero Noise :Better than 0.005 ppm RMS (1 minute average time)
    Lower Detectable Limit :0.01 ppm (1 minute average time)
    Zero Drift (24 hour) :0.005 ppm
    Span Drift (24 hour) :+/- 1% full scale
    Linearity :+/- 1% full scale
    Sample Flow Rate :Approximately 110 cc/min
    Power Requirements :90-100 VAC @ 50/60 Hz; 105-125 VAC @ 50/60 Hz; 210-250 VAC @ 50/60 Hz
    Dimensions :16.75" (W) x 8.62" (H) x 23" (D)
    Weight :53 lbs.
    Outputs :NO, NO2, NOX, selectable voltages (0-100 mV, 0-1 VDC, 0-5 VDC, 0-10 VDC), RS-232, Optional 4-20 mA, RS-485