COMPOUND
|
CO
|
CO2
|
NO
|
NO2
|
SO2
|
O2
|
|
Gas
|
Carbon Monoxide
|
Carbon Dioxide
|
Nitrogen
|
Nitrogen Dioxide
|
Sulfur Dixoide
|
Oxygen
|
|
Measurement Principle
|
Non-Dispersive Infra Red
|
Electrochemical Cell |
Paramagnetic Cell |
| Minimum Range |
0 - 100
ppm |
0 - 5% |
0 - 50
ppm |
0 - 20
ppm |
0 - 20
ppm |
0 - 5%
|
| Full Scale Range |
0 - 2,500
ppm |
0 - 25% |
0 - 2,000
ppm |
0 - 500
ppm |
0 - 10,000 ppm |
0 - 25%
|
| Lower Detection Limit |
0.5 ppm |
0.05% |
0.5 ppm |
0.2 ppm |
0.2 ppm |
0.20%
|
0.10%
|
Zero Drift
(24 Hours) |
<1 ppm |
<0.1% |
<1.2 ppm |
<1 ppm |
<0.5 ppm |
<0.2%
|
<0.1% |
Zero Drift
(7 Days) |
<3.0 ppm |
<0.5% |
<5.0 ppm |
<3.0 ppm |
<3.0 ppm |
<0.2% |
<0.1% |
Span Drift
(24 Hours) |
<1%
of span |
<1%
of span |
<1%
of span |
<1%
of span |
<1%
of span |
<0.2% |
<0.1% |
Span Drift
(7 Days) |
<1%
of span |
<1%
of span |
<1%
of span |
<1%
of span |
<1%
of span |
<0.2% |
<0.1% |
| Accuracy |
+/- 2%
of span |
+/- 2%
of span |
+/- 2%
of span |
+/- 2%
of span |
+/- 2%
of span |
+/- 0.25%
of span |
+/- 0.1%
of span |
| Response Time |
70
seconds |
70
seconds |
70 seconds |
70
seconds |
70
seconds |
60
seconds |
45
seconds |
| Linearity |
2% full scale or 5% of measured value (whichever is smaller) |
0.20% |
0.10% |
| Zero Noise |
<0.2ppm |
0.03% |
<0.2ppm |
<0.1ppm |
<0.1ppm |
0.050%
|
| Display Resolution |
0.1ppm |
0.01% |
0.1ppm |
0.1ppm |
0.1ppm |
0.1% |
0.01% |
>> To learn more about this product download a PDF (117KB)