ORP Standard Instructions

ORP readings can be reported using the absolute mV values or the EH mV values, which give the corresponding mV values that would be obtained using a Standard Hydrogen Electrode (SHE), also known as the Normal Hydrogen Electrode (NHE). ORP standard allows operators to correlate their ORP measurements to the Standard Hydrogen Electrode if they are using an Orion ORP electrode. A meter with a relative mV mode and scrolling capability is required.

 

Caution: ORP standard will stain hands and clothing. In case of contact, flush the affected area with water. A dilute sodium thiosulfate solution may help remove stains from clothing.

 

Note: Do not store ORP electrodes in ORP standard.

 

To determine ORP values as compared to the Standard Hydrogen Electrode (SHE) at 25 ˚C with Orion ORP electrodes: EH = + 420 ± 3 mV

 

1.  Place the ORP electrode in the Orion ORP standard.

2.  Set the meter to the relative mV mode.

3.  Adjust the mV reading to the EH value using Table 1.

4.  The meter is now calibrated to measure the ORP values of samples relative to the Standard Hydrogen Electrode.

 

Table 1

Temperature (°C)

EH Value (mV)

Absolute Value with Cat. No. 900011 Filling Solution (mV)

Absolute Value with Cat. No. 900001 Filling Solution (mV)

0

+438

218

176

5

+435

218

176

10

+431

219

175

15

+428

219

175

20

+424

219

174

25

+420

220

173

30

+415

220

172

35

+411

220

171

40

+406

220

170

45

+401

220

169

50

+396

220

167

Electrode Compatibilty

 

ORP standards, Cat. No. 967961 and 967901, are designed for use with following ORP combination electrodes:

 

·     Cat. No. 9778BNWP - Glass body combination ORP electrode with waterproof BNC connector

·     Cat. No. 9778SC - Glass body combination ORP electrode with screw cap connector

·     Cat. No. 9678BNWP - Sure-Flow® junction, epoxy body combination ORP electrode with waterproof BNC connector

·     Cat. No. 967800 - Sure-Flow junction, epoxy body combination ORP electrode withU.S. standard connector

·     Cat. No. 9179BNMD - Epoxy body, non-refillable gel ORP/ATC Triode® with waterproof BNC and 8 pin MiniDIN connectors

·     Cat. No. 9179BN - Epoxy body, non-refillable gel ORP/ATC Triode with BNC and 8 pin DIN connectors

·     Cat. No. 9180BNMD - Epoxy body, refillable ORP/ATC Triode with waterproof BNC and 8 pin MiniDIN connectors

·     Cat. No. 9180BN - Epoxy body, refillable ORP/ATC Triode with BNC and 8 pin DIN connectors 