|
| As there is continued interest in running AccelaTM High Speed LC as a stand alone LC (either conventional or U-HPLC), some users may be working with samples at very sensitive conditions. For these users, minimizing baseline noise is very important. In such instances, we have found a number of measures that are good operating practices and are described in more detail below: |
|
|
|
- Software
a. If using ChromQuest 4.2 ensure the Hot Fix described in FAST 2007.043 has been
installed.
2. Accela PDA Detector
a. For high speed/high pressure applications the 1 cm, 2 μL LightPipeTM Flowcell must be
used to ensure optimum peak shape.
b. Install a 100 psi back pressure regulator on the outlet tubing of the PDA detector. This
reduces the out-gassing effects eliminating the occurrence of bubbles in the flow cell
and is recommended for high pressure applications.
c. If there are already bubbles in the flowcell, then follow this procedure for "burping" the
flowcell:
i. Pump 100% methanol through the system for five minutes at 1mL/min.
ii. After the five minutes have passed and the pump is still pumping, place your
thumb over the end of the outlet tubing of the PDA detector and hold for three
seconds then release. Repeat the hold and release procedure two more times.
3. Accela Pump Pulse Dampener
a. Purge the pulse dampener because there could be a bubble in the pulse dampener
tube. The pulse dampener should be purged every 30 days and filled with Isopropyl
Alcohol (IPA).

i. Ensure that the connection from the pump outlet (3) to the autosampler
injection valve inlet is connected.
ii. Connect a syringe to the pump purge outlet (1)
iii. Connect a syringe to the pulse dampener flush (2)
iv. Open the pump purge outlet (1)
v. Start pumping 100% IPA at 1mL/min.
vi. Wait five minutes
vii. Open the pulse dampener flush (2)
viii. Close the pump purge outlet (1)
ix. Wait until air bubbles are no longer visible (at least 5 minutes)
x. Close the pulse dampener flush (2) and open the pump purge outlet (1)
xi. Wait five minutes
xii. Close the pump purge outlet (1)
4. Additional Mixing
a. For complex gradients such as those using formic acid it may be required to have
additional mixing capacity. Below are a list of mixers you can choose from:
i. 25 μL in-line static mixer P/N 00109-99-00031
ii. 50 μL in-line static mixer P/N 00109-99-00032
iii. 150 μL in-line static mixer P/N 00109-99-00033
For other great technical tips and helpful information designed to assist you in optimizing your LC methods, visit our Chromatography Resource Center.
|